Author:
Johnson Gemma,Millar Michael R,Matthews Stuart,Skyrme Margaret,Marsh Peter,Barringer Emma,O'Hara Stephen,Wilks Mark
Abstract
Abstract
Background
Methicillin-resistant Staphylococcus aureus (MRSA) is a major nosocomial pathogen worldwide. The need for accurate and rapid screening methods to detect MRSA carriers has been clearly established. The performance of a novel assay, BacLite Rapid MRSA (Acolyte Biomedica, UK) for the rapid detection (5 h) and identification of hospital associated ciprofloxacin resistant strains of MRSA directly from nasal swab specimens was compared to that obtained by culture on Mannitol salt agar containing Oxacillin (MSAO) after 48 h incubation.
Results
A total of 1382 nasal screening swabs were tested by multiple operators. The BacLite Rapid MRSA test detected 142 out of the 157 confirmed MRSA that were detected on MSAO giving a diagnostic sensitivity of 90.4, diagnostic specificity of 95.7% and a negative predictive value of 98.7%. Of the 15 false negatives obtained by the BacLite Rapid MRSA test, seven grew small amounts (< 10 colonies of MRSA) on the MSAO culture plate and five isolates were ciprofloxacin sensitive. However there were 13 confirmed BacLite MRSA positive samples, which were negative by the direct culture method, probably due to overgrowth on the MSAO plate. There were 53 false positive results obtained by the BacLite Rapid MRSA test at 5 h and 115 cases where MRSA colonies were tentatively identified on the MSAO plate when read at 48 h, and which subsequently proved not to be MRSA.
Conclusion
The Baclite MRSA test is easy to use and provides a similar level of sensitivity to conventional culture for the detection of nasal carriage of MRSA with the advantage that the results are obtained much more rapidly.
Publisher
Springer Science and Business Media LLC
Subject
Microbiology (medical),Microbiology
Cited by
27 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献