Abstract
Abstract
Background
Mammographic breast density is an important predictor of breast cancer, but its measurement has limitations related to subjectivity of visual evaluation or to difficult access for automatic volumetric measurement methods. Herein, we describe the design and clinical validation of Aguida, a software program for automated quantification of breast density from flat mammography images.
Materials and methods
The software program was developed in MatLab. After image segmentation separating the background from the breast image, the operator positions a cursor defining a region of interest on the pectoralis major muscle from the mediolateral oblique view. Then, in the craniocaudal view, the threshold for separation of the dense tissue is based on the optical density of the pectoral muscle, and the proportion of dense tissue is calculated by the program. Mammograms obtained from 2 different occasions in 291 women were used for clinical evaluation.
Results
The intraclass correlation coefficient (ICC) between breast density measurements by the software and by a radiologist was 0.96, with a bias of only 0.67 percentage points and a 95% limit of agreement of 13.5 percentage points; the ICC was 0.94 in the interobserver reliability assessment by two radiologists with different experience; and the ICC was 0.98 in the intraobserver reliability assessment. The distribution among the density classes was close to the values obtained with the volumetric software.
Conclusions
Measurement of breast density with the Aguida program from flat mammography images showed high agreement with the visual determination by radiologists, and high inter- and intra-observer reliability.
Funder
Universidade Federal do Rio Grande do Norte
Publisher
Springer Science and Business Media LLC
Subject
Health Informatics,Health Policy,Computer Science Applications
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献