Abstract
Background
The determination of the right x-ray angiography viewing angle is an important issue during the treatment of thoracic endovascular aortic repair (TEVAR). An inaccurate projection angle (manually determined today by the physicians according to their personal experience) may affect the placement of the stent and cause vascular occlusion or endoleak.
Methods
Based on the acquisition of a computed tomography angiography (CTA) image before TEVAR, an adaptive optimization algorithm is proposed to determine the optimal viewing angle of the angiogram automatically. This optimal view aims at avoiding any overlapping between the left common carotid artery and the left subclavian artery. Moreover, the proposed optimal procedure exploits the patient-specific morphology to adaptively reduce the potential foreshortening effect.
Results
Experimental results conducted on thirty-five patients demonstrate that the optimal angiographic viewing angle based on the proposed method has no significant difference when compared with the expert practice (p = 0.0678).
Conclusion
We propose a method that utilizes the CTA image acquired before TEVAR to automatically calculate the optimal C-arm angle. This method has the potential to assist surgeons during their interventional procedure by providing a shorter procedure time, less radiation exposure, and less contrast injection.
Funder
National Science and Technology Major Project of the Ministry of Science and Technology of China
National Natural Science Foundation
Publisher
Springer Science and Business Media LLC
Subject
Radiology Nuclear Medicine and imaging
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献