Author:
Huang Kemin,Feng Yanlin,Liang Weitang,Li Lin
Abstract
Abstract
Background
Image reconstruction algorithm is one of the important factors affecting the quantitative parameters of PET/CT. The purpose of this study was to investigate the effects of time of flight (TOF) and point spread function (PSF) on quantitative parameters of lung lesions in 18F-FDG PET/CT.
Methods
This retrospective study evaluated 60 lung lesions in 39 patients who had undergone 18F-fluoro-deoxy-glucose (FDG) PET/CT. All lesions larger than 10 mm in diameter were included in the study. The PET data were reconstructed with a baseline ordered-subsets expectation–maximization (OSEM) algorithm, OSEM + PSF, OSEM + TOF and OSEM + TOF + PSF respectively. The differences of maximum standard uptake value (SUVmax), mean standard uptake value (SUVmean), metabolic tumor volume (MTV), total lesion glycolysis (TLG)and signal to noise ratio (SNR)were compared among different reconstruction algorithms.
Results
Compared with OSEM reconstruction, using OSEM + TOF + PSF increased SUVmean and SUVmax by 23.73% and 22.71% respectively, and SNR increased by 70.18%, MTV decreased by 23.84% (p < 0.01). The percentage difference was significantly higher in smaller lesions (diameter 10–22 mm) than in larger lesions (diameter 23–44 mm), and significantly higher in low contrast lesions (SNR ≤ 15.31) than in high contrast lesions (SNR > 15.31). The difference of TLG among various reconstruction algorithms is relatively small, the highest value is − 6.48% of OSEM + TOF + PSF, and the lowest value is 0.81% of OSEM + TOF.
Conclusion
TOF and PSF significantly affected the quantitative parameters of lung lesions in 18F-FDG PET/CT. OSEM + TOF + PSF can significantly increased SUVmax, SUVmean and SNR, and significantly reduce MTV, especially in small lesions and low contrast lesions. TLG can be relatively stable in different reconstruction algorithms.
Publisher
Springer Science and Business Media LLC
Subject
Radiology Nuclear Medicine and imaging
Reference33 articles.
1. Beyer T, Czernin J, Freudenberg LS. Variations in clinical PET/CT operations: results of an international survey of active PET/CT users. J Nucl Med. 2011;52:303–10.
2. Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45:1519–27.
3. Nahmias C, Wahl LM. Reproducibility of standardized uptake value measurements determined by 18F-FDG PET in malignant tumors. J Nucl Med. 2008;49:1804–8.
4. Wiele C, Kruse V, Smeets P, Sathekge M, Maes A. Predictive and prognostic value of metabolic tumour volume and total lesion glycolysis in solid tumours. Eur J Nucl Med Mol Imaging. 2013;40:290–301.
5. Bazan JG, Duan F, Snyder BS, et al. Metabolic tumor volume predicts overall survival and local control in patients with stage III non-small cell lung cancer treated in ACRIN 6668/RTOG 0235. Eur J Nucl Med Mol Imaging. 2017;44:17–24.