Author:
Shi Yuan,He Deming,Hou Yingyong,Hu Qin,Xu Chen,Liu Yalan,Jiang Dongxian,Su Jieakesu,Zeng Haiying,Tan Yunshan
Abstract
Abstract
Background
Tissue microarray (TMA) is a high throughput research tool, which has greatly facilitated and accelerated in situ tissue analyses. However, its productivity has been restricted due to the confined thickness of traditional donor block. Here, we introduce an improved high output TMA method that is applicable to a broader range of tissue samples.
Methods
In this method, a 3.6 cm long and 2.7 cm wide recipient block with 88 square lattices (3 mm in width) was first prepared using several commercial instruments. A 2 mm wide and 6 mm long tissue rod was then prepared using a self-made blade-shaped knife from each paraffin embedded donor block of gastrointestinal stromal tumors. These rods were manually arrayed one by one into the corresponding lattices of the 60°C pre-softened recipient block with the guide of holes drilled with a steel needle. A 70-rod TMA was made to testify this method.
Results
The prepared TMA had well defined array configurations, good tissue morphology and fully preserved proteins and DNA. A total of 500–1000 TMA sections could be easily obtained from a TMA block.
Conclusion
This low-cost and time-saving method provides an alternative sampling tool for high output TMA.
Virtual Slides
The virtual slide(s) for this article can be found here: http://www.diagnosticpathology.diagnomx.eu/vs/1979605867857990
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Histology,Pathology and Forensic Medicine
Cited by
41 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献