Abstract
Abstract
Purpose
Highly cross-linked polyethylene has been introduced to decrease osteolysis secondary to polyethylene wear debris generation. However, few long-term data on revision total hip arthroplasty (THA) using highly cross-linked polyethylene liners are available. The objective of this study was to determine long-term outcomes of a highly cross-linked polyethylene liner in revision THA.
Materials & methods
We evaluated 63 revision THAs performed in 63 patients using a highly cross-linked polyethylene liner between April 2000 and February 2005. Of these, nine died and four were lost to follow-up. Thus, the final study cohort consisted of 50 patients (50 hips), including 26 males and 24 females with a mean age of 53 years (range, 27–75 years). Mean follow-up was 11 years (range, 10–14 years).
Results
The mean Harris hip score improved from 44 points preoperatively to 85 points at the final follow-up. No radiographic evidence of osteolysis was found in any hip. The mean rate of polyethylene liner wear was 0.029 mm/year (range, 0.003 to 0.098 mm/year). A total of 5 hips (10%) required re-revision arthroplasty, including one cup loosening, one recurrent dislocation, and three deep infections. Kaplan-Meier survivorship with an end point of re-revision for any reason was 91.1% and for aseptic cup loosening was 97.9% at 11 years.
Conclusion
At a minimum of 10 years, the highly cross-linked polyethylene liners showed excellent clinical performance and implant survivorship, and were not associated with osteolysis in our patients with revision THAs.
Publisher
Springer Science and Business Media LLC
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献