Abstract
Abstract
Objectives
Measurement of intracranial pressure (ICP) plays an important role in long-term monitoring and neuro-intensive treatment of patients with a cerebral shunt. Currently, only two complete telemetric implants with different technical features are available worldwide. This prospective pilot study aims to examine patients who had both probes implanted at overlapping times for clinical reasons and represents the first in vivo comparison of both measurement methods.
Materials and methods
Patients with a primary subarachnoid hemorrhage or a spontaneous intracerebral hemorrhage with ventricular hemorrhage who had received a telemetric ICP probe (Raumedic® NEUROVENT®-P-tel) were included in the study. Conventional external ventricular drainages (EVD) and ventriculoperitoneal shunts with a telemetric ICP probe (Miethke Sensor Reservoir) were implanted in patients with hydrocephalus who required CSF (cerebrospinal fluid) drainage. Absolute ICP values from all systems were obtained. Due to the overlapping implantation time, parallel ICP measurements were performed via two devices simultaneously. ICP measurements via the sensor reservoir were repeated after 3 and 9 months. Differences between the absolute ICP values measured via the NEUROVENT®-P-tel probe, the Miethke sensor reservoir®, and the EVD were analyzed.
Results
Seventeen patients were included in the present study between 2016 and 2018. 63% of all patients were male. In 11 patients the ICP measurements were followed up with both devices for 3 months. ICP measurements of the sensor reservoir showed corresponding trends in 9 cases compared to ICP measurement via the telemetry probe or EVD. Difference in absolute ICP values ranged between 14.5 mmHg and 0.0 mmHg. The average difference of the absolute ICP values in 8 cases was ≤ 3.5 mmHg.
Conclusion
ICP measurements with both systems continuously showed synchronous absolute ICP values, however absolute values of ICP measurement with the different systems did not match.
Publisher
Springer Science and Business Media LLC
Subject
Cellular and Molecular Neuroscience,Developmental Neuroscience,Neurology,General Medicine
Reference23 articles.
1. Antes S, Tschan CA, Kunze G, Ewert L, Zimmer A, Halfmann A, Oertel J. Clinical and radiological findings in long-term intracranial pressure monitoring. Acta Neurochir. 2014;156(5):1009–19 (discussion 19).
2. Antes S, Tschan CA, Heckelmann M, Breuskin D, Oertel J. Telemetric intracranial pressure monitoring with the Raumedic Neurovent P-tel. World Neurosurg. 2016;91:133–48.
3. Miller JD, Becker DP, Ward JD, Sullivan HG, Adams WE, Rosner MJ. Significance of intracranial hypertension in severe head injury. J Neurosurg. 1977;47(4):503–16.
4. Welschehold S, Schmalhausen E, Dodier P, Vulcu S, Oertel J, Wagner W, Tschan CA. First clinical results with a new telemetric intracranial pressure-monitoring system. Neurosurgery. 2012;70(1 Suppl Operative):44–9.
5. Kiefer M, Antes S, Leonhardt S, Schmitt M, Orakcioglu B, Sakowitz OW, Eymann R. Telemetric ICP measurement with the first CE-approved device: data from animal experiments and initial clinical experiences. Acta Neurochirurgica Suppl. 2012;114:111–6.
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献