Comparability of hemoglobin A1c level measured in capillary versus venous blood sample applying two point-of-care instruments

Author:

Keramati Tahereh,Razi Farideh,Tootee Ali,Larijani Bagher

Abstract

Abstract Background The present study is designed to evaluate the validity of the measurement of capillary blood hemoglobin A1c levels in comparison with venous blood hemoglobin A1c. Methods The data of this cross-sectional study are collected from a sample of 45 Iranian diabetic patients referred to one particular laboratory for the assessment of HbA1c level during a period from April to December 2013. Venous and simultaneous capillary blood samples were obtained from each subject for measurement of hemoglobin A1c levels. Both samples were tested using two different NGSP certified systems: CERA STAT 2000 (Ceragem Medisys Inc) and NycoCard Reader II (Axis-Shield). Results The mean hemoglobin A1c in venous and capillary blood samples measured using CERA STAT 2000 assays were 6.30 ± 1.68% and 6.34 ± 1.65% respectively (p = 0.590). However, when NycoCard Reader II assay was employed, the mean hemoglobin A1c in venous and capillary blood samples were 6.73 ± 1.35% and 6.92 ± 1.50% (p = 0.007). Moreover, a strong correlation was observed between venous and capillary hemoglobin A1c levels with Pearson’s concordance correlation coefficients of 0.96 and 0.94 with the use of NycoCard Reader II and CERA STAT 2000 assays respectively. Application of CERA STAT 2000 demonstrated to be of a considerably higher value using the ROC curve analysis assay (AUC = 0.991). Also, similar analysis by using NycoCard Reader II assay demonstrated that capillary hemoglobin A1c measurement had high value for differentiation of uncontrolled from controlled blood glucose level (AUC = 0.935). Conclusion It was demonstrated that capillary hemoglobin A1c measurement had a considerably high value for differentiating between poorly-controlled and well-controlled blood glucose levels.

Publisher

Springer Science and Business Media LLC

Subject

Endocrinology, Diabetes and Metabolism,Internal Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3