Abstract
Abstract
Background
Implantable Collamer lens (ICL) vaulting is one of the most important parameters for the safety, aqueous humor circulation, and lens transparency after ICL implantation. This study aimed to investigate the factors associated with the actual vaulting after refractive EVO-ICL surgery.
Methods
This retrospective study included patients who underwent EVO-ICL surgery at a tertiary eye hospital between October and December 2019. A RESCAN 700 was used for the intraoperative and CIRRUS HD-OCT was used for postoperative observation of vaulting. Subjective and objective refractions, anterior ocular segment, corneal morphology, intraocular pressure (IOP), anterior chamber volume (ACV), crystalline lens rise (CLR), white-to-white distance (WTW), anterior chamber depth (ACD), axial length, corneal endothelial cell density (ECD), and fundoscopy were examined. A multivariable analysis was performed to determine the factors independently associated with 1-month postoperative vaulting.
Results
Fifty-one patients (102 eyes) were included. Compared with the eyes with normal vaulting, those with high vaulting had higher preoperative diopter values (P = 0.039), lower preoperative corrected visual acuity (P = 0.006), lower preoperative IOP (P = 0.029), higher preoperative ACD (P = 0.004), lower preoperative CLR (P = 0.046), higher ICL spherical equivalent (P = 0.030), higher intraoperative vaulting (P < 0.001), and lower IOP at 1 month (P = 0.045). The multivariable analysis showed that the only factor independently associated with high vaulting at 1 month after surgery was the intraoperative vaulting value (odds ratio = 1.005, 95% confidence interval: 1.002–1.007, P < 0.001). The intraoperative and 1-month postoperative vaulting values were positively correlated (R2 = 0.562).
Conclusions
The RESCAN700 system can be used to perform intraoperative optical coherence tomography to predict the vaulting value of ICL at 1 month.
Publisher
Springer Science and Business Media LLC
Subject
Ophthalmology,General Medicine
Reference56 articles.
1. Holden BA, Wilson DA, Jong M, Sankaridurg P, Fricke TR, Smith EL III, et al. Myopia: a growing global problem with sight-threatening complications. Commun Eye Health. 2015;28(90):35.
2. Hashemi H, Fotouhi A, Yekta A, Pakzad R, Ostadimoghaddam H, Khabazkhoob M. Global and regional estimates of prevalence of refractive errors: systematic review and meta-analysis. J Current Ophthalmol. 2018;30(1):3–22.
3. Alfonso JF, Baamonde B, Fernandez-Vega L, Fernandes P, Gonzalez-Meijome JM, Montes-Mico R. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J Cataract Refract Surg. 2011;37(5):873–80.
4. Igarashi A, Shimizu K, Kamiya K. Eight-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Am J Ophthalmol. 2014;157(3):532–539 e531.
5. Bloomenstein MR, Dulaney DD, Barnet RW, Perkins SA. Posterior chamber phakic intraocular lens for moderate myopia and hyperopia. Optometry. 2002;73(7):435–46.
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献