Comparison of the new self-contained darkroom refractive screener versus table-top autorefractor and cycloplegia retinoscopy in detecting refractive error

Author:

Wei Xianxian,Li Lili,Jiang Li,Lu Haiyan,Huang Huiyao,Zhong Dedong,Pan Liang,Wei Diefeng,Han Yun,Lin Hong,Chen Qi

Abstract

Abstract Purpose By comparing the results of the new self-contained darkroom refractive screener (YD-SX-A) versus table-top autorefractor and cycloplegic retinoscopy, to evaluate the performance of the YD-SX-A in detecting refractive error in children and adolescents and then judge whether it can be used in refractive screening. Methods Cross-sectional study. 1000 participants between the ages of 6 and 18 who visited the Optometry Center of the People's Hospital of Guangxi Zhuang Autonomous Region from June to December 2022 were selected. First, participants were instructed to measure their diopter with a table-top autorefractor (Topcon KR8800) and YD-SX-A in a noncycloplegic setting. After cycloplegia, they were retinoscopy by a professional optometrist. The results measured by three methods were collected respectively. To avoid deviation, only the right eye (1000 eyes) data were used in the statistical analysis. The Bland–Altman plots were used to evaluate the agreement of diopters measured by the three methods. The receiver operating characteristic (ROC) curves was used to analysis effectiveness of detecting refractive error of YD-SX-A. Results The average age of participants was 10.77 ± 3.00 years, including 504 boys (50.4%) and 496 girls (49.6%). When YD-SX-A and cycloplegia retinoscopy (CR) were compared in the myopia group, there was no statistical difference in spherical equivalent (SE) (P > 0.05), but there was a statistical difference in diopter spherical (DS) and diopter cylinder (DC) (P < 0.05). Comparing the diopter results of Topcon KR8800 and CR, the difference between each test value in the myopia group was statistically significant (P < 0.05). In the hyperopia group, the comparison between YD-SX-A and CR showed no statistically significant differences in the DC (P > 0.05), but there were significant differences in the SE and DS (P < 0.05). In the astigmatism group, the SE, DS, and DC were statistically different, and the DC of YD-SX-A was lower than that of CR and Topcon KR8800. Bland–Altman plots indicated that YD-SX-A has a moderate agreement with CR and Topcon KR8800. The sensitivity and specificity of YD-SX-A for detecting myopia, hyperopia and astigmatism were 90.17% and 90.32%, 97.78% and 87.88%, 84.08% and 74.26%, respectively. Conclusion This study has identified that YD-SX-A has shown good performance in both agreement and effectiveness in detecting refractive error when compared with Topcon KR8800 and CR. YD-SX-A could be a useful tool for large-scale population refractive screening.

Funder

Guangxi Health Commission Self-funded Scientific Research Project

Guangxi Clinical Ophthalmic Research Center

Publisher

Springer Science and Business Media LLC

Subject

Ophthalmology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3