Author:
Li Huajin,Zheng Kailing,Wang Huihang,Xie Maosong
Abstract
Abstract
Purpose
To compare the efficacy and efficiency of self-assembled intraocular rare earth magnet and forceps in removing intraocular foreign bodies(IOFBs) undergoing 25-gauge(G) pars plana vitrectomy.
Methods
A total of 30 patients with metallic IOFB underwent 25-G PPV were enrolled into this study. Self-assembled intraocular rare earth magnet were used in 15 patients(bar group), and forceps were used in 15 patients(forceps group). Success rate of removing IOFB, time taken to remove IOFB, incidence of IOFB slippage and fall, iatrogenic retinal damages were compared between the two groups.
Results
There was no significant difference in success rate of removing IOFBs between the groups(93.3% and 100%, P > 0.99). The median time taken of removing FB was significantly shorter in bar group than in forceps group(112 and 295 s, P = 0.001). None of the patients in bar group had IOFB slippage and fall, or related iatrogenic retinal damage in the process of removal. In forceps group, IOFB slippage and fall during removal were observed in 7 of 15(47.6%) patients, related iatrogenic retinal injuries were recorded in 6 of 15(40.0%) patients, both were significantly higher than bar group(P = 0.003 and P = 0.017, respectively).
Conclusions
Compared with forceps, the assembled intraocular magnet can greatly reduce the possibility of IOFB slippage and fall, prevent related iatrogenic retinal damage, and shorten the time taken to remove IOFB. The assembled intraocular magnet can be an useful tool in removing metallic IOFBs in PPV.
Funder
Young and Middle-aged Key Personnel Training Project of Fujian Provincial Health Commission
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献