Abstract
Abstract
Background
The objective of this study was to investigate the effects of locking plugs and the biomechanical properties of a 3.5 mm 8-hole polyaxial locking plate in a fracture gap model. Our hypothesis was that locking plugs would increase the strength and stiffness of the construct. Twelve 3.5 mm 8-hole plates were used to evaluate two different construct designs (with locking plugs vs. without locking plugs) with validated bone substitutes in a 25 mm bridging osteosynthesis gap model. Each construct was subjected to a single cycle four-point bending load to failure using a servo-hydraulic testing machine. Bending stiffness, bending strength, and bending structural stiffness were calculated and compared using an unpaired Student´s t-test.
Results
The plating construct with locking plugs did not show any significant increase in terms of bending stiffness, bending strength, and bending structural stiffness compared to plating construct without locking plugs in a 25 mm gap fracture model during a single cycle four-point bending.
Conclusions
Under the conditions tested, filling empty plate holes with locking plugs in bridging osteosynthesis does not increase stiffness or strength of the plate-bone construct.