Lapsina Sandra,Riond Barbara,Hofmann-Lehmann Regina,Stirn Martina
Abstract
Abstract
Background
Bronchoalveolar lavage (BAL) is a diagnostic method for the assessment of the lower respiratory airway health status in horses. Differential cell count and sometimes also total nucleated cell count (TNCC) are routinely measured by time-consuming manual methods, while faster automated methods exist. The aims of this study were to compare: 1) the Sysmex XN-V body fluid (BF) mode with the manual techniques for TNCC and two-part differential into mononuclear and polymorphonuclear cells; 2) the Olympus VS200 slide scanner and software generated deep-learning-based algorithm with manual techniques for four-part differential cell count into alveolar macrophages, lymphocytes, neutrophils, and mast cells. The methods were compared in 69 clinical BAL samples.
Results
Incorrect gating by the Sysmex BF mode was observed on many scattergrams, therefore all samples were reanalyzed with manually set gates. For the TNCC, a proportional and systematic bias with a correlation of r = 0.79 was seen when comparing the Sysmex BF mode with manual methods. For the two-part differential count, a mild constant and proportional bias and a very small mean difference with moderate limits of agreement with a correlation of r = 0.84 and 0.83 were seen when comparing the Sysmex BF mode with manual methods. The Sysmex BF mode classified significantly more samples as abnormal based on the TNCC and the two-part differential compared to the manual method. When comparing the Olympus VS200 deep-learning-based algorithm with manual methods for the four-part differential cell count, a very small bias in the regression analysis and a very small mean difference in the difference plot, as well as a correlation of r = 0.85 to 0.92 were observed for all four cell categories. The Olympus VS200 deep-learning-based algorithm also showed better precision than manual methods for the four-part differential cell count, especially with an increasing number of analyzed cells.
Conclusions
The Sysmex XN-V BF mode can be used for TNCC and two-part differential count measurements after reanalyzing the samples with manually set gates. The Olympus VS200 deep-learning-based algorithm correlates well with the manual methods, while showing better precision and can be used for a four-part differential cell count.
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献