Accuracy of the end-expiratory lung volume measured by the modified nitrogen washout/washin technique: a bench study

Author:

Berger-Estilita JoanaORCID,Haenggi Matthias,Ott Daniel,Berger David

Abstract

Abstract Background The functional residual capacity (FRC) determines the oxygenating capacity of the lung and is heavily affected in the clinical context of the acute respiratory distress syndrome. Nitrogen-wash-in/wash-out methods have been used to measure FRC. These methods have rarely been validated against exactly known volumes. The aim of the study was to assess the accuracy and precision of the N2 washout/washin method in measuring FRC, by comparing it with set volumes in a lung simulator. Methods We conducted a diagnostic bench study in the Intensive Care Unit and Radiology Department of a tertiary hospital in Switzerland. Using a fully controllable high fidelity lung simulator (TestChest®), we set the functional residual capacity at 1500 ml, 2000 ml and 2500 ml and connected to the GE Carestation respirator, which includes the nitrogen washout/washin technique (INview™ tool). FRC was then set to vary by different levels of PEEP (5, 8, 12 and 15 cmH2O). The main outcome measures were bias and precision of the TestChest® when compared to the results from the washout/washin technique, according to the results of a Bland Altman Analysis. We verified our findings with volumetric computed tomography. Results One hundred and thirty-five nitrogen-wash-in/wash-out measurements were taken at three levels of FIO2 (0.4, 0.5, 0.6). The CT volumetry reproduced the set end-expiratory volumes at the Simulator with a bias of 4 ml. The nitrogen-wash-in/wash-out method had a bias of 603 ml with acceptable limits of agreement (95% CI 252 to − 953 ml). Changes were detected with a concordance rate of 97%. Conclusions We conclude that the TestChest® simulator is an accurate simulation tool, concerning the simulation of lung volumes. The nitrogen wash-in/wash out method correlated positively with FRC changes, despite a relatively large bias in absolute measurements. The reference volumes in the lung simulator verified with CT volumetry were very close to their expected values. The reason for the bias could not be determined.

Publisher

Springer Science and Business Media LLC

Subject

General Biochemistry, Genetics and Molecular Biology,General Medicine

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3