Abstract
Abstract
Background
An impedance threshold device (ITD) was developed to increase venous return to the heart and therefore increase cardiac output and organ blood flow during cardiopulmonary rescue (CPR). Basic CPR aims to maintain coronary and cerebral blood flow at the minimum level necessary for survival. The present study compared the effects of an ITD on cerebral blood flow assessed as blood flow in both carotid arteries to the blood flow of a control group during prolonged CPR.
Methods
Fourteen anaesthetized pigs were monitored during 60 min of CPR after induced ventricular fibrillation. The primary outcome was blood flow in both carotid arteries, and the secondary outcomes were blood pressure, acid–base parameters, plasma potassium, and plasma lactate. The pigs were randomized to mechanical compressions and ventilation with an ITD added to the ventilation or to a control group treated only with mechanical compressions and ventilation. The time course for the parameters was tested using analysis of variance.
Results
The cumulative carotid blood flow in the ITD group decreased from 64 to 42 ml/min, and it decreased from 69 to 51 ml/min in the control group during 60 min of CPR. The difference was not significant. The secondary outcome measures were also not significantly different.
Conclusions
This study did not show any beneficial effect of an ITD on carotid blood flow.
Publisher
Springer Science and Business Media LLC
Subject
General Biochemistry, Genetics and Molecular Biology,General Medicine
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献