Modern physiological approach to inappropriate ICD shocks due to atrial fibrillation with very fast ventricular response. A case report
-
Published:2024-03-12
Issue:1
Volume:24
Page:
-
ISSN:1471-2318
-
Container-title:BMC Geriatrics
-
language:en
-
Short-container-title:BMC Geriatr
Author:
Pestrea CatalinORCID, Enache RoxanaORCID, Cicala EcaterinaORCID, Vatasescu RaduORCID
Abstract
Abstract
Background
Fast-conducting atrial fibrillation misinterpreted as ventricular tachycardia is the leading cause for inappropriate shocks in patients with implantable cardiac defibrillators (ICD). These inappropriate shocks are associated with significant morbidity and mortality and cause great discomfort and stress.
Case presentation
We report the case of a patient with ischemic cardiomyopathy, permanent atrial fibrillation, and a single-chamber DF-1 ICD implanted for the primary prevention of sudden cardiac death, who presented for multiple inappropriate internal shocks due to very fast-conducting atrial fibrillation, which was mislabeled as ventricular fibrillation by the ICD. Since the patient was under maximal atrioventricular nodal blocking medical therapy (beta-blockers and digitalis) and we didn`t find any reversible causes for the heart rate acceleration, we opted for rate control with atrioventricular node ablation. To counteract the risk of pacing-induced cardiomyopathy in this patient who would become totally pacemaker-dependent, we successfully performed left bundle branch area pacing. Because the patient`s ICD had a DF-1 connection and the battery had a long life remaining, we connected the physiological pacing lead to the IS-1 sense-pace port of the ICD. The 6-month follow-up showed an improvement in left ventricular function with no more inappropriate shocks.
Conclusions
Left bundle branch area pacing and atrioventricular node ablation in patients with an implantable single-chamber DF-1 defibrillator and fast-conducting permanent atrial fibrillation is a cost-efficient and very effective method to prevent and treat inappropriate shocks, avoiding the use of an additional dual-chamber or CRT-D device.
Publisher
Springer Science and Business Media LLC
Reference10 articles.
1. van Rees JB, Borleffs CJ, de Bie MK, Stijnen T, van Erven L, Bax JJ, Schalij MJ. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality. J Am Coll Cardiol. 2011;57(5):556– 62. doi: 10.1016/j.jacc.2010.06.059. PMID: 21272746. 2. Groenveld HF, Tijssen JG, Crijns HJ, Van den Berg MP, Hillege HL, Alings M, Van Veldhuisen DJ, Van Gelder IC. RACE II Investigators. Rate control efficacy in permanent atrial fibrillation: successful and failed strict rate control against a background of lenient rate control: data from RACE II (Rate Control Efficacy in Permanent Atrial Fibrillation). J Am Coll Cardiol. 2013;61(7):741-8. https://doi.org/10.1016/j.jacc.2012.11.038. PMID: 23410544. 3. Brignole M, Menozzi C, Gianfranchi L, Musso G, Mureddu R, Bottoni N, Lolli G. Assessment of atrioventricular junction ablation and VVIR pacemaker versus pharmacological treatment in patients with heart failure and chronic atrial fibrillation: a randomized, controlled study. Circulation. 1998;98(10):953– 60. https://doi.org/10.1161/01.cir.98.10.953. PMID: 9737514. 4. Kiehl EL, Makki T, Kumar R, Gumber D, Kwon DH, Rickard JW, Kanj M, Wazni OM, Saliba WI, Varma N, Wilkoff BL, Cantillon DJ. Incidence and predictors of right ventricular pacing-induced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic function. Heart Rhythm. 2016;13(12):2272–2278. https://doi.org/10.1016/j.hrthm.2016.09.027. PMID: 27855853. 5. Chung MK, Patton KK, Lau CP, Dal Forno ARJ, Al-Khatib SM, Arora V, Birgersdotter-Green UM, Cha YM, Chung EH, Cronin EM, Curtis AB, Cygankiewicz I, Dandamudi G, Dubin AM, Ensch DP, Glotzer TV, Gold MR, Goldberger ZD, Gopinathannair R, Gorodeski EZ, Gutierrez A, Guzman JC, Huang W, Imrey PB, Indik JH, Karim S, Karpawich PP, Khaykin Y, Kiehl EL, Kron J, Kutyifa V, Link MS, Marine JE, Mullens W, Park SJ, Parkash R, Patete MF, Pathak RK, Perona CA, Rickard J, Schoenfeld MH, Seow SC, Shen WK, Shoda M, Singh JP, Slotwiner DJ, Sridhar ARM, Srivatsa UN, Stecker EC, Tanawuttiwat T, Tang WHW, Tapias CA, Tracy CM, Upadhyay GA, Varma N, Vernooy K, Vijayaraman P, Worsnick SA, Zareba W, Zeitler EP. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heart Rhythm. 2023;20(9):e17–91. https://doi.org/10.1016/j.hrthm.2023.03.1538. Epub 2023 May 20. PMID: 37283271.
|
|