Accuracy and trending abilities of finger plethysmographic blood pressure and cardiac output compared to invasive measurements during caesarean delivery in healthy women: an observational study

Author:

Omenås Ivar N.,Tronstad Christian,Rosseland Leiv ArneORCID

Abstract

Abstract Background In women presenting for caesarean section under spinal anesthesia, continuous measurement of circulatory aspects, such as blood pressure and cardiac output, is often needed. At present, invasive techniques are used almost exclusively. Reliable non-invasive monitoring would be welcome, as it could be safer, less uncomfortable, and quick and easy to apply. We aimed to evaluate whether a non-invasive, finger plethysmographic device, the ccNexFin monitor, can replace invasively measured blood pressure in the radial artery, and whether cardiac output measurements from this device can be used interchangeably with measurements from the mini-invasive LiDCO monitor currently in use at our institution. Methods Simultaneous invasive measurements were compared to ccNexFin in 23 healthy women during elective caesarean section under spinal anesthesia. We used Bland Altman statistics to assess agreement, and polar plot methodology to judge trending abilities with pre-defined limits. Results Mean arterial and systolic pressures showed biases (invasive – ccNexFin) of − 4.3 and 12.2 mmHg, with limits of agreement of − 15.9 – 7.4 and − 11.1 – 35.6, respectively. The ccNexFin trending abilities were within the suggested limits for mean pressure but insufficient for systolic pressure compared to invasive measurements. Cardiac output had a small bias of 0.2 L/min, but wide limits of agreement of − 2.6 – 3.0. The ccNexFin trending abilities compared to the invasive estimated values (LiDCO) were unsatisfactory. Conclusions We consider the ccNexFin monitor to have sufficient accuracy in measuring mean arterial pressure. The limits of agreement for systolic measurements were wider, and the trending ability compared to invasive measurements was outside the recommended limit. The ccNexFin is not reliable for cardiac output measurements or trend in pregnant women for caesarean delivery under spinal anesthesia. Trial registration Registered May 23, 2013, at ClinicalTrials.gov under number NCT01861132.

Publisher

Springer Science and Business Media LLC

Subject

Anesthesiology and Pain Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3