Double lumen endobronchial tubes (DLTs) are frequently used to employ single lung ventilation strategies during thoracic surgical procedures. Placement of these tubes can be challenging even for experienced clinicians. We hypothesized that airway anatomy, particularly of the glottis and proximal trachea, significantly impacts the ease or difficulty in placement of these tubes.
Methods
Images from 24 randomly selected Positron Emission Tomography – Computed Tomography (PET-CT) scans were evaluated for several anatomic aspects of the upper airway, including size and angulation of the glottis and proximal tracheal using calibrated CT measurements and an online digital protractor. The anatomic issues identified were confirmed in cadaveric anatomic models.
Results
Proximal tracheal diameter measurements in PET-CT scans demonstrated a mean ± standard deviation of 20.4 ± 2.5 mm in 12 males and 15.5 ± 0.98 mm in 12 females (p < 0.001), and both were large enough to accommodate 39 French and 37 French DLTs in males and females, respectively. Subsequent measurements of the posterior angulation of the proximal trachea revealed a mean angle of 40.8 ± 5.7 degrees with no sex differences. By combining the 24 individual posterior tracheal angles with the 16 angled distal tip measurements DLTs (mean angle 24.9 ± 2.1 degrees), we created a series of 384 patient intubation angle scenarios. This data clearly showed that DLT rotation to a full 180 degrees decreased the mean intubation angle between the DLT and the proximal trachea from a mean of 66.6 ± 5.9 to only 15.8 ± 5.9 degrees.
Conclusions
Rotation of DLTs a full 180 instead of the recommended 90 degrees facilitates DLT intubations.
Publisher
Springer Science and Business Media LLC
Reference13 articles.
1. Pedersen CM, Green JS, Bigler DR, Andersen NE, Cromhout PF. Evaluation of time to intubation and rate of success for different healthcare professionals using a double-lumen left-sided endotracheal video tube: a prospective observational study. J Perioper Pract. 2020;30(12):383–8. PMID: 32096441.
2. Pollard JE, Smith DW, Morgan DE, Skaggs JD. Combined use of a total control introducer and a hyperangulated video laryngoscope to place a left-sided double lumen endotracheal tube in a patient with a history of difficult laryngoscopy. BMJ Case Rep. 2021;14(10):e243407. PMID: 34625438.
3. De Marchi L, Patel J, Razmjou K. Lung Ultrasound in Thoracic Surgery: Confirming Placement of a Pediatric Right Double-Lumen Tube. A A Pract. 2020;14(10):e01296. PMID: 32845107.
4. Huang P, Zhou R, Lu Z, Hang Y, Wang S, Huang Z. GlideScope® versus C-MAC®(D) videolaryngoscope versus Macintosh laryngoscope for double lumen endotracheal intubation in patients with predicted normal airways: a randomized, controlled, prospective trial. BMC Anesthesiol. 2020;20(1):119. PMID: 32434470.
5. Muncey AR, Sweeney J, Hafez O. Placement of a double-lumen endotracheal tube using a combined bronchoscopy and fluoroscopy technique: a Case Report. Pract. 2020;14(6):e01198. PMID: 32784316.