Abstract
Abstract
Background
Recent studies indicate that ultrasound can detect changes in tracheal diameter during endotracheal tube (ETT) cuff inflation. We sought to assess the accuracy of ultrasound measurement of tracheal diameter, and to determine the relationship between tracheal wall pressure (TWP), cuff inflation volume (CIV), and the degree of tracheal deformation.
Methods
Our study comprised two parts: the first included 45 porcine tracheas, the second 41 porcine tracheas. Each trachea was intubated with a cuffed ETT, which was connected to an injector and the manometer via a three-way tap. The cuff was inflated and the cuff pressure recorded before and after intubation. The tracheal diameter was measured using ultrasound. This included three separate measurements: outer transverse diameter (OTD), internal transverse diameter (ITD), and anterior tracheal wall thicknesses (ATWT). A precision electronic Vernier caliper was also used to measure tracheal diameter. We calculated TWP and the percentage change of tracheal diameter. The Bland–Altman method, linear regression, and locally weighted regression (LOESS) were used to analyze the data.
Results
There were strong correlation and agreement for OTD (r = 0.97, P < 0.001) and ITD (r = 0.90, P < 0.001) as measured by ultrasound and by precision electronic Vernier caliper, but a poor correlation for ATWT (r = 0.58, P < 0.001). There was a strong correlation between the percentage change of OTD (OTD%, r = 0.75, P < 0.001) and CIV, the percentage change of ITD (ITD%, r = 0.77, P < 0.001) and CIV, TWP (r = 0.75, P < 0.001) and CIV. And a strong correlation was also found between TWP and OTD% (r = 0.84, P < 0.001), TWP and ITD% (r = 0.84, P < 0.001).
Conclusions
Use of ultrasound to measure OTD and ITD is accurate, but is less accurate for ATWT. There is a close correlation between OTD%, ITD%, CIV and TWP.
Funder
the National Key Research and Development Program of China
Publisher
Springer Science and Business Media LLC
Subject
Anesthesiology and Pain Medicine
Reference29 articles.
1. You-Ten KE, Siddiqui N, Teoh WH, Kristensen MS. Point-of-care ultrasound (POCUS) of the upper airway. Can J Anaesth. 2018;65:473–84.
2. Mandeep S, Ki Jinn C, Chan VWS, Wong DT, Prasad GA, Eugene Y. Use of sonography for airway assessment: an observational study. J Ultrasound Med Official J Am Inst Ultrasound Med. 2010;29:79.
3. Werner SL, Smith CE, Goldstein JR, Jones RA, Cydulka RK. Pilot study to evaluate the accuracy of ultrasonography in confirming endotracheal tube placement. Ann Emerg Med. 2007;49:75–80.
4. Bunyamin M, Hüseyin S, Ahmet K, Rüveyda Irem D, Muhammet GZ, Burhanettin U, et al. Use of sonography for rapid identification of esophageal and tracheal intubations in adult patients. J Ultrasound Med. 2011;30:671–6.
5. Hoffmann B, Gullett JP, Hill HF, Fuller D, Westergaard MC, Hosek WT, et al. Bedside ultrasound of the neck confirms endotracheal tube position in emergency intubations. Ultraschall Med. 2014;35:451–8.