Author:
Eley Victoria,Christensen Rebecca,Guy Louis,Wyssusek Kerstin,Pelecanos Anita,Dodd Benjamin,Stowasser Michael,van Zundert Andre
Abstract
Abstract
Background
Measuring blood pressure in patients with obesity is challenging. The ClearSight™ finger cuff (FC) uses the vascular unloading technique to provide continuous non-invasive blood pressure measurements. We aimed to test the agreement of the FC with invasive radial arterial monitoring (INV) in patients with obesity.
Methods
Participants had a body mass index (BMI) ≥45 kg/m2 and underwent laparoscopic bariatric surgery. FC and INV measurements were obtained simultaneously every 5 min on each patient, following induction of anesthesia. Agreement over time was assessed using modified Bland-Altman plots and error grid analysis permitted clinical interpretation of the results. Four-quadrant plots allowed assessment of concordance in blood pressure changes.
Results
The 30 participants had a median (IQR) BMI of 50.2 kg/m2 (IQR 48.3–55.3). The observed bias (SD, 95% limits of agreement) for systolic blood pressure (SBP) was 14.3 mmHg (14.1, -13.4 – 42.0), 5.2 mmHg (10.9, -16.0 – 26.5) for mean arterial pressure (MAP) and 2.6 mmHg (10.8, -18.6 – 23.8) for diastolic blood pressure (DBP). Error grid analysis showed that the proportion of readings in risk zones A-E were 90.8, 6.5, 2.7, 0 and 0% for SBP and 91.4, 4.3, 4.3, 0 and 0% for MAP, respectively. Discordance occurred in ≤8% of pairs for consecutive change in SBP, MAP and DBP.
Conclusions
The vascular unloading technique was not adequately in agreement with radial arterial monitoring. Evaluation in a larger sample is required before recommending this technique for intraoperative monitoring of patients with BMI ≥45 kg/m2.
Publisher
Springer Science and Business Media LLC
Subject
Anesthesiology and Pain Medicine
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献