Comparison of post-auricular and frontal bispectral index values obtained during renal surgeries

Author:

Abdelrahman Ahmed Mohammed FetouhORCID,Elbadry Amr Arafa,Omara Amany Faheem

Abstract

Abstract Background The bispectral index (BIS) monitor is one of the EEG-derived monitoring techniques and well-established devices used to measure the depth of anesthesia. This study aimed to assess the agreement of BIS values based on the positions of either post-auricular or frontal sensors in individual patients undergoing renal surgeries while lateral positions at various stages of anesthesia. Patients and methods 12 patients older than 18 years, ASA I-III patients scheduled for elective renal operations, two BIS were placed on each patient, one on each side of the post-auricular region and one across the forehead, and each sensor was connected to a different BIS monitor. We gathered three pieces of data at each of the six-time points: BIS score, signal quality index (SQI) score calculating the signal’s strength and electromyography (EMG) score: before the onset of anesthesia (awake) when the eyelash reflex is lost (LOC), after intubation (intubation), following the initial surgical incision, each 30 min throughout the procedure (maintenance), and at the moment the patient’s eyes open naturally after waking up from anesthesia (emergence). Results The overall BIS value at the frontal position was significantly higher than the post-auricular position (52.5 ± 22.2 and 52.1 ± 22.1, respectively, P = 0.010). On the other hand, the BIS value was comparable between the frontal and post-auricular positions at LOC, intubation, 60, 120, and 80 min and at emergence. A strong link between the two sensor positions, as indicated by the correlation coefficient (r = 0.607, P < 0.001), and the Bland-Altman analysis revealed a small mean difference (-1.8) and a low (9.0/- 12.5) limit of agreement, with just 4.3% of the readings falling outside of it during the anesthetic maintenance period. Conclusion Acceptable variation in BIS data was observed when obtained from the two different sensor positions for clinical usage. The post-auricular BIS sensor system may be a suitable substitute for an impractical frontal setup. Protocol Registration The study was registered in clinicaltrials.gov on 11/07/2022 (trial registration number: NCT05451823).

Funder

Tanta University

Publisher

Springer Science and Business Media LLC

Subject

Anesthesiology and Pain Medicine

Reference22 articles.

1. Hima C, Asheeta A, Nair CC. Nair SMJWJoPS: a review on brainwave therapy. 2020:59–66.

2. Oliveira CRD, Bernardo WM, Nunes VM. Benefit of general anesthesia monitored by bispectral index compared with monitoring guided only by clinical parameters. Systematic review and meta-analysis. Revista brasileira de anestesiologia. 2017;67:72–84.

3. Hasan DR, Hassan WMNW, Hassan MH, Zaini RHM, Omar SC. A comparison of sustainable recording between cerebral state index (CSI) and bispectral index (BIS) monitors during total intravenous anaesthesia using target-controlled infusion technique for elective supratentorial craniotomy. Anaesth Pain Intensive Care. 2021;25(6):763–70.

4. Akavipat P, Hungsawanich N, Jansin R. Alternative placement of bispectral index electrode for monitoring depth of anesthesia during neurosurgery. Acta Med Okayama. 2014;68(3):151–5.

5. Nelson P, Nelson JA, Chen AJ, Kofke WA. An alternative position for the BIS-Vista montage in frontal approach neurosurgical cases. J Neurosurg Anesthesiol. 2013;25(2):135–42.