The efficacy of apneic oxygenation during intubation using a prototype of an oxygenation laryngoscope - a technical simulation

Author:

Wetsch Wolfgang A,Schroeder Daniel C,Herff Susanne J,Böttiger Bernd W,Wenzel Volker,Herff Holger

Abstract

Abstract Background Recently, a non-commercial oxygenation laryngoscope was able to maintain apneic oxygenation during simulated intubation efforts. Since that prototype was 3 mm wider than a standard Macintosh laryngoscope blade, the intubation performance of this device may differ from standard blades. A new prototype of an oxygenation laryngoscope was developed, consisting of a standard-size Macintosh blade and a fixed oxygen supply line to the side. Actually, it is unclear at which point of this blade the oxygen supply line should end to facilitate the best possible oxygen supply for apneic oxygenation. Methods In this simulation study using a standardized human airway manikin, the efficacy of apneic oxygenation by oxygen insufflation using standard and modified Macintosh blades was compared: a standard Macintosh blade without oxygen supply line as control, one with an additional oxygen supply line ending proximal near the handle, one with the line ending at the middle of the blade, and one with the line ending near the tip. A preoxygenated test lung was connected to an oximeter with a flow rate of 200ml/min, simulating oxygen consumption of a male adult, and to the trachea of an anatomically correctly shaped airway manikin. Apneic oxygenation was performed and oxygen content was measured over a 20-minutes observation period. Experiments were repeated five times for each laryngoscope blade. Results Oxygen percentage in the test lung dropped from 100 ± 0% at the start of the experiment to 53 ± 1.5% in the room air control group (p < 0.001 compared to all other groups), and to 74 ± 2.5% in the proximal oxygen line group, whereas oxygen percentage remained at 100% in both the medium and distal oxygen line groups (p = 1 between these groups; p < 0.001 between all other groups). Conclusions In this simulation study with a preoxygenated airway manikin, the use of a modified Macintosh laryngoscope blade with oxygen line attached at the tip or at the middle were able to maintain apneic oxygenation without measurable drop of oxygen content over 20 min. Proximal placement of the oxygen supply line still showed an advantage against room air, however it did not completely prevent room air from entering the airway. Trial registration Not applicable.

Funder

Universitätsklinikum Köln

Publisher

Springer Science and Business Media LLC

Subject

Anesthesiology and Pain Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3