Abstract
Abstract
Background
Biological diagnosis of hemoglobin disorders is a complex process relying on the combination of several analytical techniques to identify Hb variants in a particular sample. Currently, hematology laboratories usually use high-performance liquid chromatography (HPLC), capillary electrophoresis and gel-based methods to characterize Hb variants. Co-elution and co-migration may represent major issues for precise identification of Hb variants, even for the most common ones such as Hb S and C.
Methods
We adapted a top-down selected reaction monitoring (SRM) electron transfer dissociation (ETD) mass spectrometry (MS) method to fit with a clinical laboratory environment. An automated analytical process with semi-automated data analysis compatible with a clinical practice was developed. A comparative study between a reference HPLC method and the MS assay was performed on 152 patient samples.
Results
The developed workflow allowed to identify with high specificity and selectivity the most common Hb variants (Hb S and Hb C). Concordance of the MS-based approach with HPLC was 71/71 (100%) for Hb S and 11/11 (100%) for Hb C.
Conclusions
This top-down SRM ETD method can be used in a clinical environment to detect Hb S and Hb C.
Funder
Swiss National foundation
Fondation Ernst et Lucie Schmidheiny
Publisher
Springer Science and Business Media LLC
Subject
Clinical Biochemistry,Molecular Biology,Molecular Medicine,Clinical Biochemistry,Molecular Biology,Molecular Medicine
Cited by
3 articles.
订阅此论文施引文献