Abstract
Abstract
Background
This study evaluated the operating performance of an implant navigation system used by dental students and dentists of prosthodontic background with varying levels of experience. A surgical navigation system and optical tracking system were used, and dentists’ accuracies were evaluated in terms of differences between the positions of actually drilled holes and those of the holes planned using software before surgeries.
Methods
The study participants were 5 dental students or dentists who had studied in the same university and hospital but had different experience levels regarding implants. All participants were trained in operating the AqNavi system in the beginning of the study. Subsequently, using 5 pairs of dental models, each participant drilled 5 implant holes at 6 partially edentulous positions (11, 17, 26, 31, 36, and 47). In total, each participant conducted 30 drilling tests.
Results
In total, 150 tests among 5 dentists at 6 tooth positions (11, 17, 26, 31, 36, and 47) were conducted. Although a comparison of the tests revealed significant differences in the longitudinal error (P < .0001) and angular error (P = .0011), no significant difference was observed in the total error among the dentists.
Conclusions
A relatively long operating time was associated with relatively little implant experience. Through the dental navigation system, dental students can be introduced to dental implant surgery earlier than what was possible in the past. The results demonstrated that the operational accuracy of the dental implant navigation system is not restricted by participants’ implant experience levels. The implant navigation system assists the dentist in the ability to accurately insert the dental implant into the correct position without being affected by his/her own experience of implant surgery.
Funder
Southern Taiwan Science Park Bureau, Ministry of Science and Technology
Ministry of Science and Technology
Publisher
Springer Science and Business Media LLC
Reference23 articles.
1. Kobayashi K, Shimoda S, Nakagawa Y, Yamamoto A. Accuracy in measurement of distance using limited cone-beam computerized tomography. Int J Oral Maxillofac Implants. 2004;19:228–31.
2. Sammartino G, Marenzi G, di Lauro AE, Paolantoni G. Aesthetics in oral implantology: biological, clinical, surgical, and prosthetic aspects. Implant Dent. 2007;16:54–65.
3. Ewers R, Schicho K, Truppe M, Seemann R, Reichwein A, Figl M, Wagner A. Computer- aided navigation in dental implantology: 7 years of clinical experience. Journal of oral and maxillofacial surgery. J Oral Maxillofac Surg. 2004;62:329–34.
4. Birkfellner W, Solar P, Gahleitner A, Huber K, Kainberger F, Kettenbach J, et al. In-vitro assessment of a registration protocol for image guided implant dentistry. Clin Oral Implants Res. 2001;12:69–78.
5. Sießegger M, Schneider BT, Mischkowski RA, Lazar F, Krug B, Klesper B, et al. Use of an image-guided navigation system in dental implant surgery in anatomically complex operation sites. J Craniomaxillofac Surg. 2001;29:276–81.
Cited by
42 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献