Author:
Liu Chaowei,Xing Yifeng,Li Yan,Lin Yanjun,Xu Jianghan,Wu Dong
Abstract
Abstract
Introduction
The aim of this study was to verify whether the use of short implants could optimize stress distribution of bone surrounding implants in atrophic mandibles with different bone qualities.
Methods
A three-dimensional model of the atrophic mandible with three levels of bone quality was made using computer software. Short implants (6 mm) and standard implants (10 mm) were used in four designs: Design 1 "All-On four", Design 2 "All-On-four" with two short implants, Design 3 four vertical implants with two short implants, and Design 4 six short implants. The distal short implants were placed at the first molar position. All twelve models were imported into finite element analysis software, and 110 N oblique force was loaded on the left second premolar. Maximum principal stress values of peri-implant bone and the volumes of bone with over 3000 microstrians (overload)were analyzed.
Result
Stress values and volumes of overload bone increased in all four groups with the decline of bone quality. The highest stress values were found in the cortical bone surrounding the Design 1 inclined implant in two lower bone quality mandibles, and the lowest in Design 3. However, Design 1 had less overload bone tissue than all three designs with short implants.
Conclusion
Short implants placed posteriorly helped decrease stress values in peri-implant bone, while bone surrounding short implants had a high resorption risk in low bone quality mandible.
Funder
Startup Fund for scientific research, Fujian Medical University
Medical Innovation Project of Fujian Province
Natural Science Foundation of Fujian Province
Publisher
Springer Science and Business Media LLC
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献