Background
Attachment material is one of the contributing factors to the degree of wear of the attachment components in mandibular implant-retained overdentures. The purpose of this in vitro study was to compare the wear behavior of 2 different attachment systems of dissimilar materials in mandibular implant-retained overdentures by qualitative and quantitative methods.
Methods
Two attachment systems of different materials were utilized (n = 16); Titach (Dental Evolutions Inc, Beverly Hills, CA, USA) with a titanium-to-titanium interface and Locator R-Tx (Zest Anchors Inc, Escondido, CA, USA) with a titanium-to-nylon interface. One thousand cycles of overdenture insertion and removal simulating 1-year clinical use were performed. All matrices were removed from the overdentures and all patrices were unscrewed from the implants for wear assessment quantitively using a stereomicroscope and qualitatively using a scanning electron microscope. Data were analyzed by using an independent sample t test.
Results
After cyclic loading, stereomicroscopic findings showed that the Titach group had statistically significant higher wear value than the Locator R-Tx group (p < 0.001). Moreover, scanning electron microscopy images showed noticeable abrasion in Titach patrix at the area of highest convexity. However, the Locator R-Tx matrix displayed an apparent tear of rubber inserts.
Conclusions
Titach attachment with the titanium-to-titanium interface revealed more wear than Locator R-Tx attachment with the titanium-to-nylon interface. Thus, the type of attachment material influences the degree of wear of the attachment components.
Publisher
Springer Science and Business Media LLC
Reference38 articles.
1. Mahanna FF, Elsyad MA, Mourad SI, Abozaed HW. Satisfaction and Oral Health-Related Quality of Life of Different Attachments Used for Implant-Retained Overdentures in Subjects with Resorbed Mandibles: A Crossover Trial. Int J Oral Maxillofac Implants. 2020;35(2):423–31.
2. Awad MA, Rashid F, Feine JS, Overdenture Effectiveness Study Team Consortium. The effect of mandibular 2-implant overdentures on oral health-related quality of life: an international multicentre study. Clin Oral Implants Res. 2014;25(1):46–51.
3. Emami E, Heydecke G, Rompré PH, de Grandmont P, Feine JS. Impact of implant support for mandibular dentures on satisfaction, oral and general health-related quality of life: a meta-analysis of randomized-controlled trials. Clin Oral Implants Res. 2009;20(6):533–44.
4. Feine JS, Carlsson GE, Awad MA, Chehade A, Duncan WJ, Gizani S, et al. The McGill consensus statement on overdentures: mandibular two-implant overdentures as first choice standard of care for edentulous patients. Int J Oral Maxillofac Implants. 2002;17(4):601–2.
5. Thomason JM, Feine J, Exley C, Moynihan P, Müller F, Naert I, et al. Mandibular two implant-supported overdentures as the first choice standard of care for edentulous patients–the York Consensus Statement. Br Dent J. 2009;207(4):185–6.