Abstract
Background
Extracorporeal membrane oxygenation (ECMO), an invasive mechanical therapy, provides cardio-respiratory support to critically ill patients when maximal conventional support has failed. ECMO is delivered via large-bore cannulae which must be effectively secured to avoid complications including cannula migration, dislodgement and accidental decannulation. Growing evidence suggests tissue adhesive (TA) may be a practical and safe method to secure vascular access devices, but little evidence exists pertaining to securement of ECMO cannulae. The aim of this study was to determine the safety and efficacy of two TA formulations (2-octyl cyanoacrylate and n-butyl-2-octyl cyanoacrylate) for use in peripherally inserted ECMO cannula securement, and compare TA securement to ‘standard’ securement methods.
Methods
This in vitro project assessed: (1) the tensile strength and flexibility of TA formulations compared to ‘standard’ ECMO cannula securement using a porcine skin model, and (2) the chemical resistance of the polyurethane ECMO cannulae to TA. An Instron 5567 Universal Testing System was used for strength testing in both experiments.
Results
Securement with sutures and n-butyl-2-octyl cyanoacrylate both significantly increased the force required to dislodge the cannula compared to a transparent polyurethane dressing (p = 0.006 and p = 0.003, respectively) and 2-octyl cyanoacrylate (p = 0.023 and p = 0.013, respectively). Suture securement provided increased flexibility compared to TA securement (p < 0.0001), and there was no statistically significant difference in flexibility between 2-octyl cyanoacrylate and n-butyl-2-octyl cyanoacrylate (p = 0.774). The resistance strength of cannula polyurethane was not weakened after exposure to either TA formulation after 60 min compared to control.
Conclusions
Tissue adhesive appears to be a promising adjunct method of ECMO cannula insertion site securement. Tissue adhesive securement with n-butyl-2-octyl cyanoacrylate may provide comparable securement strength to a single polypropylene drain stitch, and, when used as an adjunct securement method, may minimise the risks associated with suture securement. However, further clinical research is still needed in this area.
Funder
The Prince Charles Hospital Foundation
Publisher
Springer Science and Business Media LLC
Subject
Critical Care and Intensive Care Medicine
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献