Efficacy of endotracheal tube clamping to prevent positive airways pressure loss and pressure behavior after reconnection: a bench study

Author:

Bulleri Enrico,Fusi Cristian,Bambi Stefano,Pisani Luigi,Galesi Alice,Rizzello Enrico,Lucchini Alberto,Merlani Paolo,Pagnamenta AlbertoORCID

Abstract

Abstract Background Endotracheal tube (ETT) clamping before disconnecting the patient from the mechanical ventilator is routinely performed in patients with acute respiratory distress syndrome (ARDS) to minimize alveolar de-recruitment. Clinical data on the effects of ETT clamping are lacking, and bench data are sparse. We aimed to evaluate the effects of three different types of clamps applied to ETTs of different sizes at different clamping moments during the respiratory cycle and in addition to assess pressure behavior following reconnection to the ventilator after a clamping maneuver. Methods A mechanical ventilator was connected to an ASL 5000 lung simulator using an ARDS simulated condition. Airway pressures and lung volumes were measured at three time points (5 s, 15 s and 30 s) after disconnection from the ventilator with different clamps (Klemmer, Chest-Tube and ECMO) on different ETT sizes (internal diameter of 6, 7 and 8 mm) at different clamping moments (end-expiration, end-inspiration and end-inspiration with tidal volume halved). In addition, we recorded airway pressures after reconnection to the ventilator. Pressures and volumes were compared among different clamps, different ETT-sizes and the different moments of clamp during the respiratory cycle. Results The efficacy of clamping depended on the type of clamp, the duration of clamping, the size of the ETT and the clamping moment. With an ETT ID 6 mm all clamps showed similar pressure and volume results. With an ETT ID 7 and 8 mm only the ECMO clamp was effective in maintaining stable pressure and volume in the respiratory system during disconnection at all observation times. Clamping with Klemmer and Chest-Tube at end inspiration and at end inspiration with halved tidal volume was more efficient than clamping at end expiration (p < 0.03). After reconnection to the mechanical ventilator, end-inspiratory clamping generated higher alveolar pressures as compared with end-inspiratory clamping with halved tidal volume (p < 0.001). Conclusions ECMO was the most effective in preventing significant airway pressure and volume loss independently from tube size and clamp duration. Our findings support the use of ECMO clamp and clamping at end-expiration. ETT clamping at end-inspiration with tidal volume halved could minimize the risk of generating high alveolar pressures following reconnection to the ventilator and loss of airway pressure under PEEP.

Funder

University of Geneva

Publisher

Springer Science and Business Media LLC

Subject

Critical Care and Intensive Care Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3