Abstract
Abstract
Objectives
Acoustic noise in magnetic resonance imaging (MRI) negatively impacts patients. We assessed a silent gradient coil switched at 20 kHz combined with a T1-weighted magnetisation prepared rapid gradient-echo (MPRAGE) sequence at 7 T.
Methods
Five healthy subjects (21–29 years; three females) without previous 7-T MRI experience underwent both a quiet MPRAGE (Q-MPRAGE) and conventional MPRAGE (C-MPRAGE) sequence twice. Image quality was assessed quantitatively, and qualitatively by two neuroradiologists. Sound level was measured objectively and rated subjectively on a 0 to 10 scale by all subjects immediately following each sequence and after the whole examination (delayed). All subjects also reported comfort level, overall experience and willingness to undergo the sequence again.
Results
Compared to C-MPRAGE, Q-MPRAGE showed higher signal-to-noise ratio (10%; p = 0.012) and lower contrast-to-noise ratio (20%; p < 0.001) as well as acceptable to good image quality. Q-MPRAGE produced 27 dB lower sound level (76 versus 103 dB). Subjects reported lower sound level for Q-MPRAGE both immediate (4.4 ± 1.4 versus 6.4 ± 1.3; p = 0.007) and delayed (4.6 ± 1.4 versus 6.3 ± 1.3; p = 0.005), while they rated comfort level (7.4 ± 1.0 versus 6.1 ± 1.7; p = 0.016) and overall experience (7.6 ± 1.0 versus 6.0 ± 0.9; p = 0.005) higher. Willingness to undergo the sequence again was also higher, however not significantly (8.1 ± 1.0 versus 7.2 ± 1.3; p = 0.066).
Conclusion
Q-MPRAGE using a silent gradient coil reduced sound level by 27 dB compared to C-MPRAGE at 7 T while featuring acceptable-to-good image quality and a quieter and more pleasant subject experience.
Publisher
Springer Science and Business Media LLC
Subject
Radiology, Nuclear Medicine and imaging
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献