Monitoring heparin therapy: stability of two different anti-Xa assays using blood samples collected in citrate-containing and CTAD tubes

Author:

Gremillet Marion,Talon Laurie,Lebreton Aurélien,Sinegre Thomas

Abstract

Abstract Background Anti-factor Xa assays and activated partial thromboplastin time (aPTT) are mainly employed to monitor patients treated with heparins. According to the Clinical and Laboratory Standards Institute and the French Working Group on Haemostasis and Thrombosis, anti-factor Xa activity and aPTT should be tested within 2 h of blood sampling for unfractionated heparin (UFH) monitoring. However, discrepancies exist depending on the used reagents and collecting tubes. The study aim was to determine the stability of aPTT and anti-factor Xa measurements using blood samples collected in citrate-containing or citrate-theophylline-adenosine-dipyridamole (CTAD) tubes and stored for up to 6 h. Methods Patients receiving UFH or low molecular weight heparin (LMWH) were enrolled; aPTT and anti-factor Xa activity were tested using two different analyser/reagent pairs (Stago and reagent without dextran sulfate; Siemens and reagent with dextran sulfate) after 1, 4 and 6 h of sample storage as whole blood or as plasma. Results For UFH monitoring, comparable anti-factor Xa activity and aPTT results were obtained with both analyser/reagent pairs when samples were stored as whole blood before plasma isolation. With samples stored as plasma, anti-factor Xa activity and aPTT were not affected up to 6 h after sampling when using the Stago/no-dextran sulfate reagent pair. With the Siemens/dextran sulfate-containing reagent, aPTT was significantly altered after 4 h of storage. For LMWH monitoring, anti-factor Xa activity remained stable (whole blood and plasma) for at least 6 h. Results were comparable with citrate-containing and CTAD tubes. Conclusions Anti-factor Xa activity in samples stored as whole blood or plasma was stable for up to 6 h, regardless of the reagent (with/without dextran sulfate)/collection tube. Conversely, aPTT was more variable because other plasma parameters can influence its measure and complicate the interpretation of its variations after 4 h.

Funder

Centre Hospitalier Universitaire de Clermont-Ferrand

Publisher

Springer Science and Business Media LLC

Subject

Hematology

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3