Catheter navigation by intracardiac echocardiography enables zero-fluoroscopy linear lesion formation and bidirectional cavotricuspid isthmus block in patients with typical atrial flutter

Author:

Luani Blerim,Basho Maksim,Ismail Ammar,Rauwolf Thomas,Kaese Sven,Tobli Ndricim,Samol Alexander,Pankraz Katharina,Schmeisser Alexander,Wiemer Marcus,Braun-Dullaeus Rüdiger C.,Genz Conrad

Abstract

Abstract Introduction One of the most helpful aspects of intracardiac echocardiography (ICE) implementation in electrophysiological studies (EPS) is the real-time visualisation of catheters and cardiac structures. In this prospective study, we investigated ICE-guided zero-fluoroscopy catheter navigation during radiofrequency (RF) ablation of the cavotricuspid isthmus (CTI) in patients with typical atrial flutter (AFL). Methods and results Thirty consecutive patients (mean age 72.9 ± 11.4 years, 23 male) with ongoing (n = 23) or recent CTI-dependent AFL underwent an EPS, solely utilizing ICE for catheter navigation. Zero-fluoroscopy EPS could be successfully accomplished in all patients. Mean EPS duration was 41.4 ± 19.9 min, and mean ablation procedure duration was 20.8 ± 17.1 min. RF ablation was applied for 6.0 ± 3.1 min (50W, irrigated RF ablation). Echocardiographic parameters, such as CTI length, prominence of the Eustachian ridge (ER), and depth of the CTI pouch on the ablation plane, were assessed to analyse their correlation with EPS- or ablation procedure duration. The CTI pouch was shallower in patients with an ablation procedure duration above the median (4.8 ± 1.1 mm vs. 6.4 ± 0.9 mm, p = 0.04), suggesting a more lateral ablation plane in these patients, where the CTI musculature is stronger. CTI length or ER prominence above the respective median did not correlate with longer EPS duration. Conclusions Zero-fluoroscopy CTI ablation guided solely by intracardiac echocardiography in patients with CTI-dependent AFL is feasible and safe. ICE visualisation may help to localise the optimal ablation plane, detect and correct poor tissue contact of the catheter tip, and recognise early potential complications during the ablation procedure. Graphical Abstract

Publisher

Springer Science and Business Media LLC

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3