Abstract
AbstractPhotoacoustic (PA) imaging enables noninvasive volumetric imaging of biological tissues by capturing the endogenous optical absorption contrast. Conventional ultrasound detectors using piezoelectric materials have been widely used for transducing ultrasound signals into the electrical signals for PA imaging reconstruction. However, their inherent limitations in detection bandwidth and sensitivity per unit area have unfortunately constrained the performance of PA imaging. Optical based ultrasound detection methods emerge to offer very promising solutions. In particular, polymer micro-ring resonators (MRRs) in the form of integrated photonic circuits (IPC) enable significant reduction for the sensing area to 80 μm in diameter, while maintaining highly sensitive ultrasound detection with noise equivalent pressure (NEP) of 0.49 Pa and a broad detection frequency range up to 250 MHz. The continued engineering innovation has further transformed MRRs to be transparent to the light and thus, opens up a wide range of applications, including multi-modality optical microscope with isometric resolution, PA endoscope, photoacoustic computed tomography (PACT), and more. This review article summarizes and discusses the evolution of polymer MRR design and the associated nanofabrication process for improving the performance of ultrasound detection. The resulting novel imaging applications will also be reviewed and discussed.
Funder
National Institute of General Medical Sciences
Publisher
Springer Science and Business Media LLC
Subject
General Engineering,General Materials Science
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献