Author:
Ahadi Golnaz,Welch Christian S,Grimm Michele J,Fisher David J,Zadicario Eyal,Ernström Karin,Voie Arne H,Hölscher Thilo
Abstract
Abstract
Background
The primary goal of this study was to investigate the relationship between increasing output power levels and clot fragmentation during high-intensity focused ultrasound (HIFU)-induced thrombolysis.
Methods
A HIFU headsystem, designed for brain applications in humans, was used for this project. A human calvarium was mounted inside the water-filled hemispheric transducer. Artificial thrombi were placed inside the skull and located at the natural focus point of the transducer. Clots were exposed to a range of acoustic output power levels from 0 to 400 W. The other HIFU operating parameters remained constant. To assess clot fragmentation, three filters of different mesh pore sizes were used. To assess sonothrombolysis efficacy, the clot weight loss was measured.
Results
No evidence of increasing clot fragmentation was found with increasing acoustic intensities in the majority of the study groups of less than 400 W. Increasing clot lysis could be observed with increasing acoustic output powers.
Conclusion
Transcranial sonothrombolysis could be achieved in vitro within seconds in the absence of tPA and without producing relevant clot fragmentation, using acoustic output powers of <400 W.
Publisher
Springer Science and Business Media LLC
Subject
Radiology Nuclear Medicine and imaging
Reference29 articles.
1. Dumont TM, Mokin M, Sorkin GC, Levy EI: Siddiqui AH. 2013, Journal of Neurointerventional Surgery: Aspiration thrombectomy in concert with stent thrombectomy, doi:10.1136/neurintsurg-2012-010624.rep.
2. Marder VJ, Chute DJ, Starkman S, Abolian AM, Kidwell C, Liebeskind D, Ovbiagele B, Vinuela F, Duckwiler G, Jahan R, Vespa PM, Selco S, Rajajee V, Kim D, Sanossian N, Saver JL: Analysis of thrombi retrieved from cerebral arteries of patients with acute ischemic stroke. Stroke. 2006, 37 (8): 2086-93. 10.1161/01.STR.0000230307.03438.94. doi:10.1161/01.STR.0000230307.03438.94.
3. Prokop AF, Soltani A, Roy RA: Cavitational mechanisms in ultrasound-accelerated fibrinolysis. Ultrasound Med Biol. 2007, 33 (6): 924-33. 10.1016/j.ultrasmedbio.2006.11.022.
4. Braaten JV, Goss RA, Francis CW: Ultrasound reversibly disaggregates fibrin fibers. Thromb Haemost. 1997, 78 (3): 1063-68.
5. Devcic-Kuhar B, Pfaffenberger S, Gherardini L, Mayer C, Groschl M, Kaun C, Benes E, Tschachler E, Huber K, Maurer G, Wojta J, Gottsauner-Wolf M: Ultrasound affects distribution of plasminogen and tissue-type plasminogen activator in whole blood clots in vitro. Thromb Haemost. 2004, 92 (5): 980-85.
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献