Author:
Li Qiao-yuan,Dai Wen-long,Lin can-can,Liu Xu,Guo Cheng-jun,Jian-zeng Dong
Abstract
Abstract
Background
Congenitally corrected transposition of the great arteries (ccTGA) is a rare cardiac anomaly and can lead to abnormal electrical activity of the heart. The implant of a pacemaker in such patients is more complicated than conventional operations. This case report of an adult with ccTGA who had a leadless pacemaker implant will provide a reference for diagnosing and treating such patients.
Case presentation
A 50-year-old male patient was admitted to hospital having experienced intermittent vision loss for a month. An electrocardiogram and Holter monitoring showed intermittent third-degree atrioventricular block, and echocardiography, cardiac computed tomography and cardiac magnetic resonance imaging confirmed a diagnosis of ccTGA. A leadless pacemaker was successfully implanted into the patient’s anatomical left ventricle, and the postoperative parameters were stable.
Conclusion
Implanting a leadless pacemaker into a patient with a rare anatomical and electrophysiological abnormality, such as ccTGA, is feasible and efficacious, but preoperative imaging evaluation is of considerable importance.
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,General Medicine,Surgery,Pulmonary and Respiratory Medicine
Reference5 articles.
1. Kutty S, Danford DA, Diller GP, Tutarel O. Contemporary management and outcomes in congenitally corrected transposition of the great arteries. Heart. 2018;104(14):1148–55.
2. Baruteau AE, Abrams DJ, Ho SY, Thambo JB, McLeod CJ, Shah MJ. Cardiac conduction system in congenitally corrected transposition of the great arteries and its clinical relevance. J Am Heart Assoc. 2017;6(12):e007759.
3. Yoon N, Kim KH, Park HW, et al. Complete atrioventricular block in an adult with congenitally corrected transposition of the great arteries with interrupted inferior vena cava. J Res Med Sci. 2012;17:808.
4. Shah RS, Davidson WR Jr, Samii SM. Leadless pacemaker implantation in a subpulmonic left ventricle in a patient with congenitally corrected transposition of the great arteries. HeartRhythm Case Rep. 2022;8(7):471–4.
5. Baumgartner H, De Backer J, Babu-Narayan SV, et al. 2020 ESC Guidelines for the management of adult congenital heart disease. Eur Heart J. 2021;42(6):563–645.