Abstract
Abstract
Background
Malformations of the temporal bone present different challenges to the implantation of a transcutaneous active bone conduction device, such as Bonebridge (Med-el, Innsbruck, Austria). This study aims to describe the benefits of high-resolution computed tomography (HRCT) in preoperative assessment and to analyze whether characteristics of the mastoid process, intraoperative compression of the dura or sigmoid sinus, and the use of the Lifts system, lead to differences in audiological performance after implantation.
Methods
We examined 110 cases of congenital microtia. The structure of the temporal bone was examined using HRCT and a 3D simulation software program. The mean anteroposterior mastoid bone thickness from the external auditory canal to the sigmoid sinus was measured (a measurement referred to as “AP”, hereafter). Sound field threshold (SFT), speech reception threshold (SRT) in noise, and word recognition score (WRS) in quiet, before and after implantation, were also measured. Independent variables were recorded in all patients: mastoid type (well pneumatized or poorly pneumatized), the presence of dural or sigmoid sinus compression, and the use of the Lifts system.
Results
We found that the mean AP in the non-compression group was 16.2 ± 2.3 mm and in the compression group, 13.1 ± 2.9 mm (p < 0.001). We analyzed the hearing improvement of patients grouped by mastoid development, dural or sigmoid sinus compression, and use of the Lifts system, and found that these factors did not interact and that they had no influence on the hearing outcomes (p > 0.05).
Conclusions
The AP dimension in the non-compression group was significantly larger than that in the compression group. This finding combined with the ROC curve analysis revealed the AP dimension was a high-accuracy predictor of potential surgical complications involving the dura and sigmoid sinus compression. Further analysis revealed that there was no interaction between the chosen variables: mastoid type, dural or sigmoid sinus compression, and the use of the Lifts system, and that all of these factors had no significant impact on hearing performance. Bonebridge was shown to produce effective and stable bone conduction and to improve patients’ hearing performance.
Funder
the General Programs of Beijing Municipal Science and Technology Commission
Publisher
Springer Science and Business Media LLC
Subject
Otorhinolaryngology,Surgery
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献