Slobod Douglas,Spinelli Elena,Crotti Stefania,Lissoni Alfredo,Galazzi Alessandro,Grasselli Giacomo,Mauri Tommaso
Abstract
Abstract
Background
Optimal noninvasive respiratory support for patients with hypoxemic respiratory failure should minimize work of breathing without increasing the transpulmonary pressure. Recently, an asymmetrical high flow nasal cannula (HFNC) interface (Duet, Fisher & Paykel Healthcare Ltd), in which the caliber of each nasal prong is different, was approved for clinical use. This system might reduce work of breathing by lowering minute ventilation and improving respiratory mechanics.
Methods
We enrolled 10 patients ≥ 18 years of age who were admitted to the Ospedale Maggiore Policlinico ICU in Milan, Italy, and had a PaO2/FiO2 < 300 mmHg during HFNC support with a conventional cannula. We investigated whether the asymmetrical interface, compared to a conventional high flow nasal cannula, reduces minute ventilation and work of breathing. Each patient underwent support with the asymmetrical interface and the conventional interface, applied in a randomized sequence. Each interface was provided at a flow rate of 40 l/min followed by 60 l/min. Patients were continuously monitored with esophageal manometry and electrical impedance tomography.
Results
Application of the asymmetrical interface resulted in a −13.5 [−19.4 to (−4.5)] % change in minute ventilation at a flow rate of 40 l/min, p = 0.006 and a −19.6 [−28.0 to (−7.5)] % change at 60 l/min, p = 0.002, that occurred despite no change in PaCO2 (35 [33–42] versus 35 [33–43] mmHg at 40 l/min and 35 [32–41] versus 36 [32–43] mmHg at 60 l/min). Correspondingly, the asymmetrical interface lowered the inspiratory esophageal pressure–time product from 163 [118–210] to 140 [84–159] (cmH2O*s)/min at a flow rate of 40 l/min, p = 0.02 and from 142 [123–178] to 117 [90–137] (cmH2O*s)/min at a flow rate of 60 l/min, p = 0.04.
The asymmetrical cannula did not have any impact on oxygenation, the dorsal fraction of ventilation, dynamic lung compliance, or end-expiratory lung impedance, suggesting no major effect on PEEP, lung mechanics, or alveolar recruitment.
Conclusions
An asymmetrical HFNC interface reduces minute ventilation and work of breathing in patients with mild-to-moderate hypoxemic respiratory failure supported with a conventional interface. This appears to be primarily driven by increased ventilatory efficiency due to enhanced CO2 clearance from the upper airway.
Funder
Università degli Studi di Milano
Publisher
Springer Science and Business Media LLC
Subject
Critical Care and Intensive Care Medicine
Reference28 articles.
1. Rochwerg B, Einav S, Chaudhuri D, Mancebo J, Mauri T, Helviz Y, et al. The role for high flow nasal cannula as a respiratory support strategy in adults: a clinical practice guideline. Intensive Care Med. 2020;46(12):2226–37.
2. Grieco DL, Maggiore SM, Roca O, Spinelli E, Patel BK, Thille AW, et al. Non-invasive ventilatory support and high-flow nasal oxygen as first-line treatment of acute hypoxemic respiratory failure and ARDS. Intensive Care Med. 2021;47(8):851–66.
3. Mauri T, Wang YM, Dalla Corte F, Corcione N, Spinelli E, Pesenti A. Nasal high flow: physiology, efficacy and safety in the acute care setting, a narrative review. Open Access Emerg Med. 2019;11:109–20.
4. Rochwerg B, Granton D, Wang DX, Helviz Y, Einav S, Frat JP, et al. High flow nasal cannula compared with conventional oxygen therapy for acute hypoxemic respiratory failure: a systematic review and meta-analysis. Intensive Care Med. 2019;45(5):563–72.
5. Mauri T, Turrini C, Eronia N, Grasselli G, Volta CA, Bellani G, et al. Physiologic effects of high-flow nasal cannula in acute hypoxemic respiratory failure. Am J Respir Crit Care Med. 2017;195(9):1207–15.
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献