Recurrent implant fractures in total hip arthroplasty: a yes-you-yan case report

Author:

JELSMA J,ACHARYA M

Abstract

This case report describes a single patient with recurrent implant fractures of his left total hip replacement. According to our knowledge this is the first patient in literature with recurrent implant fractures. This is a rare phenomenon as reason for revision. Risk factors for implant failure of total hip replacement include a lack of proximal support, a distally well fixed stem with proximal debonding, malalignment of the stem and raised BMI

Publisher

Universa BV

Reference10 articles.

1. van Doesburg PG, van Langelaan EJ, Apachitei I, Bénard MR, Verdegaal SHM. Femoral prosthesis neck fracture following total hip arthroplasty - a systematic review. Arthroplasty. 2020 Oct 16;2(1):28.

2. National Joint Registry for England and Wales. https://reports.njrcentre.org.uk/downloads: p. Accessed 12 April 2022.

3. Sadoghi P, Pawelka W, Liebensteiner MC, Williams A, Leithner A, Labek G. The incidence of implant fractures after total hip arthroplasty. Int Orthop. 2014 Jan;38(1):39-46.

4. Cankaya D, Yoldas B, Yılmaz S, Tecirli A, Ozkurt B. Fracture of the cemented femoral component following hemiarthroplasty in physically active patient: A case report. J Orthop Case Rep. 2015 Oct-Dec;5(4):21-3.

5. Jarvi K and Kerry RM. Segmental stem fracture of a cemented femoral prosthesis. J Arthroplasty. 2007 Jun;22(4):612-6.