Abstract

To improve the quality of prosthetic vision, it is important to understand how retinal neurons respond to electric stimulation. Previous studies present conflicting reports as to the maximum rate at which retinal ganglion cells can “follow” pulse trains, i.e., generate one spike for each pulse of the train. In the present study, we measured the response of 5 different types of rabbit retinal ganglion cells to pulse trains of 100–700 Hz. Surprisingly, we found significant heterogeneity in the ability of different types to follow pulse trains. For example, brisk transient (BT) ganglion cells could reliably follow pulse rates up to 600 pulses per second (PPS). In contrast, other types could not even follow rates of 200 PPS. There was additional heterogeneity in the response patterns across those types that could not follow high-rate trains. For example, some types generated action potentials in response to approximately every other pulse, whereas other types generated one spike per pulse for a few consecutive pulses and then did not generate any spikes in response to the next few pulses. Interestingly, in the types that could not follow high-rate trains, we found a second type of response: many pulses of the train elicited a biphasic waveform with an amplitude much smaller than that of standard action potentials. This small waveform was often observed following every pulse for which a standard spike was not elicited. A possible origin of the small waveform and its implication for effective retinal stimulation are discussed.

Publisher

American Physiological Society

Subject

Physiology,General Neuroscience

Cited by 44 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3