Gas exchange in healthy rabbits during high-frequency oscillatory ventilation

Author:

Boynton B. R.1,Hammond M. D.1,Fredberg J. J.1,Buckley B. G.1,Villanueva D.1,Frantz I. D.1

Affiliation:

1. Department of Pediatrics, Tufts University School of Medicine, Boston, Massachusetts.

Abstract

We examined the effects of oscillatory frequency (f), tidal volume (VT), and mean airway pressure (Paw) on respiratory gas exchange during high-frequency oscillatory ventilation of healthy anesthetized rabbits. Frequencies from 3 to 30 Hz, VT from 0.4 to 2.0 ml/kg body wt (approximately 20–100% of dead space volume), and Paw from 5 to 20 cmH2O were studied. As expected, both arterial partial pressure of O2 and CO2 (PaO2 and PaCO2, respectively) were found to be related to f and VT. Changing Paw had little effect on blood gas tensions. Similar values of PaO2 and PaCO2 were obtained at many different combinations of f and VT. These relationships collapsed onto a single curve when blood gas tensions were plotted as functions of f multiplied by the square of VT (f. VT2). Simultaneous tracheal and alveolar gas samples showed that the gradient for PO2 and PCO2 increased as f. VT2 decreased, indicating alveolar hypoventilation. However, venous admixture also increased as f. VT2 decreased, suggesting that ventilation-perfusion inequality must also have increased.

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

Cited by 57 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3