Limitations of a pulsed Doppler velocimeter for blood flow measurement in small vessels

Author:

Quail A. W.1,Cottee D. B.1,White S. W.1

Affiliation:

1. Discipline of Human Physiology and Neuroscience Group, Faculty of Medicine, University of Newcastle, New South Wales, Australia.

Abstract

The performance of a new and simplified flow probe construction and the Iowa 545C-4 pulsed Doppler velocimeter was evaluated for measurement of blood flow over several months in small arteries of awake animals. Calibrations were performed over a wide range of intraluminal pressures and physiological flow velocities. Pressure-dependent differences in slope of the Doppler shift-volume flow relationship were detected in some probes. Signal strength was maintained at hematocrits > 10%. Distortion of pulsed Doppler signal peaks occurred in the conscious rabbit at peak aortic velocities, at which Reynold's number for turbulence was exceeded and the Doppler shift surpassed the Nyquist limit of 31.25 kHz for the velocimeter. Although the Doppler shift-volume flow relationship is linear at < 5 kHz, in some cases at higher Doppler shifts and blood flow velocities the relationship may become nonlinear, thus causing the volume flow rate to be underestimated by up to 38%. The cause of this phenomenon may be "aliasing" and/or the consequence of the range control capability of the velocimeter selectively sampling changing velocity profiles and flow disturbances in the central stream at higher velocities.

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3