Author:
Abdul-Rasool I. H.,Chamberlain J. H.,Swan P. C.,Mitchell F. T.
Abstract
This paper reports a new system for the continuous measurements of respiratory gas exchange in ventilated subjects. It involves mixing some of the inspired gas with all of the expired gas and withdrawing the mixture at a constant rate through a dry gas meter that measures the flow. The inspired gas and expired gas mixtures are sampled and O2 and CO2 concentrations measured with a paramagnetic gas analyzer and a capnograph, respectively, to an accuracy of 0.01%. Evidence is presented to confirm the necessary stability and sensitivity of these instruments. It is possible to use the system with high inspired O2 concentrations, with ventilators where there is incomplete separation of inspired and expired gas, and in the presence of intermittent mandatory ventilation, positive end-expiratory pressure, and continuous airway pressure. The system was compared with the N2-dilution method and with the collection of expired gas in a Douglas bag in dog experiments and with patients in the intensive therapy unit. Excellent correlation between these methods was found in all circumstances.
Publisher
American Physiological Society
Subject
Physiology (medical),Physiology