Affiliation:
1. Biomedical Engineering Program, Graduate School of Engineering, Federal University of Rio de Janeiro, Rio de Janeiro, Brazil, 21945-970
Abstract
Giannella-Neto, A., C. Bellido, R. B. Barbosa, and M. F. Vidal Melo. Design and calibration of unicapillary pneumotachographs. J. Appl. Physiol.84(1): 335–343, 1998.—This study presents a method for design and calibration of unicapillary pneumotachographs for small-animal experiments. The design, based on Poiseuille’s law, defines a set of internal radius and length values that allows for laminar flow, measurable pressure differences, and minimal interference with animal’s respiratory mechanics and gas exchange. A third-order polynomial calibration (Pol) of the pressure-flow relationship was employed and compared with linear calibration (Lin). Tests were done for conditions of ambient pressure (Pam) and positive pressure (Ppos) ventilation at different flow ranges. A physical model designed to match normal and low compliance in rats was used. At normal compliance, Pol provided lower errors than Lin for mixed (1–12 ml/s), mean (4–10 ml/s), and high (8–12 ml/s) flow rate calibrations for both Pam and Ppos inspiratory tests (P < 0.001 for all conditions) and expiratory tests ( P < 0.001 for all conditions). At low compliance, they differed significantly with 8.6 ± 4.1% underestimation when Lin at Pam was used in Ppos tests. Ppos calibration, preferably in combination with Pol, should be used in this case to minimize errors (Pol = 0.8 ± 0.5%, Lin = 6.5 ± 4.0%, P < 0.0005). Nonlinear calibration may be useful for improvement of flow and volume measurements in small animals during both Pam and Ppos ventilation.
Publisher
American Physiological Society
Subject
Physiology (medical),Physiology
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献