Author:
Lloyd T. C.,Juratsch C. E.
Abstract
The Laks catheter is a triple-lumen balloon catheter used to distend the canine main pulmonary artery while recording right ventricular pressure and the arterial pressure distal to the balloon. A rise in arterial pressure reported to occur during distension has been attributed to vasoconstriction rather than passive obstruction by the balloon. We tested this in six anesthetized dogs by inflating the Laks catheter-balloon while recording pressure distal to the balloon from the Laks catheter as well as from additional catheters in right and left pulmonary arteries placed retrogradely through lobar branches following thoracotomy. We found that balloon inflation increased pressures in the arterial port of the Laks catheter and in the left pulmonary artery catheter but reduced it in the right pulmonary artery. Tightening a snare around the right pulmonary artery had the same effects on pressures. Similar results were obtained while cardiac output was controlled by left ventricular bypass perfusion in four dogs. We conclude that the Laks catheter-balloon obstructs flow to the right lung and that the arterial pressure rise recorded in it during balloon inflation cannot be distinguished from that caused by occlusion of the right pulmonary artery.
Publisher
American Physiological Society
Subject
Physiology (medical),Physiology
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献