Evaluation of a hot-wire respiratory flowmeter for clinical applicability

Author:

Yoshiya I.,Shimada Y.,Tanaka K.

Abstract

A hot-wire flowmeter was evaluated for clinical applicability. 1) Calibration with a syringe could be done if emptying duration was 1-15 s. 2) Outputs linearly decreased with decreasing concentration of N2 in O2. Outputs with CO2 in O2 and N2O in O2 represented nonlinear convex relationships with varying concentrations of CO2 and N2O. Correction for each gas mixture to be measured is required. 3) Outputs linearly decreased with decreasing barometric pressure. 4) Stability assessed with a piston respirator was excellent (within +/- 2% of syringe volume) after 15 min warmup time. However, daily calibrations are recommended in clinical situations. 5) Nebulization, if not excessive, was acceptable if the expired gas was measured at the mouth. 6) Hot-wire burning, which occurred when it was partially in contact with materials whose specific heat differed with air, was successfully protected with a simple shutoff circuit. 7) The possibility of producing nitrogen oxides by the catalytic action of the platinum hot-wire was denied by colorimetric determination. Interchangeability and sterilizability of transducers and improved mechanical strength with platinum-rhodium alloy are also discussed.

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

Cited by 31 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3