Author:
Rodenstein D. O.,Francis C.,Stanescu D. C.
Abstract
Exercise Physiol. 52: 638-641, 1982) have shown in dogs that airway closure may induce rib cage deformation and nonhomogeneous alveolar pressure swings, and they have suggested that this could lead to thoracic gas volume (TGV) overestimation by body plethysmography. However, in humans the rib cage is less easy to distort than in dogs. In four healthy volunteers we measured TGV by plethysmography before (B) and during (D) the occlusion of the middle and lower right lobes by a balloon (attached to a double-lumen catheter) positioned in the intermediate right bronchus. Subjects were trained to perform panting maneuvers preferentially with intercostals and accessory muscles or the diaphragm. Five to eleven TGV measurements were made in each subject with each panting pattern B and D occlusion. Balloon inflation resulted in no change in TGV whether low [13.3 +/- 3.4 (SD) cmH2O] or high (46.8 +/- 8.4 cmH2O) transdiaphragmatic pressures (Pdi) were used: TGV 4.0 +/- 0.4 (B) vs. 4.0 +/- 0.4 liters (D) and 4.3 +/- 0.4 (B) vs. 4.3 +/- 0.4 liters (D) for low and high Pdi, respectively. Thus, in trained subjects performing maneuvers aimed to distort the rib cage, no pressure difference was observed between the occluded and the nonoccluded lung during panting against the closed shutter. We conclude that it is unlikely that the mechanism proposed by Brown et al. might explain errors in lung volume measurements by body plethysmography in humans.
Publisher
American Physiological Society
Subject
Physiology (medical),Physiology
Cited by
14 articles.