Flow time through esophagogastric junction derived during high-resolution impedance-manometry studies: a novel parameter for assessing esophageal bolus transit

Author:

Lin Zhiyue1,Imam Hala2,Nicodème Frèdèric3,Carlson Dustin A.1,Lin Chen-Yuan1,Yim Brandon1,Kahrilas Peter J.1,Pandolfino John E.1

Affiliation:

1. Department of Medicine, Feinberg School of Medicine, Northwestern University, Chicago, Illinois;

2. Gastroenterology and Hepatology Unit, Department of Internal Medicine, Assiut University Hospital, Assiut, Egypt

3. Department of Thoracic Surgery, Centre Hospitalier de l'Université de Montréal, Montreal, Quebec, Canada; and

Abstract

This study aimed to develop and validate a method to measure bolus flow time (BFT) through the esophagogastric junction (EGJ) using a high-resolution impedance-manometry (HRIM) sleeve. Ten healthy subjects were studied with concurrent HRIM and videofluoroscopy; another 15 controls were studied with HRIM alone. HRIM studies were performed using a 4.2-mm-outer diameter assembly with 36 pressure sensors at 1-cm intervals and 18 impedance segments at 2-cm intervals (Given Imaging, Los Angeles, CA). HRIM and fluoroscopic data from four barium swallows, two in the supine and two in the upright position, were analyzed to create a customized MATLAB program to calculate BFT using a HRIM sleeve comprising three sensors positioned at the crural diaphragm. Bolus transit through the EGJ measured during blinded review of fluoroscopy was almost identical to BFT calculated with the HRIM sleeve, with the nadir impedance deflection point used as the signature of bolus presence. Good correlation existed between videofluoroscopy for measurement of upper sphincter relaxation to beginning of flow [ R = 0.97, P < 0.001 (supine) and R = 0.77, P < 0.01 (upright)] and time to end of flow [ R = 0.95, P < 0.001 (supine) and R = 0.82, P < 0.01 (upright)]. The medians and interquartile ranges (IQR) of flow time though the EGJ in 15 healthy subjects calculated using the virtual sleeve were 3.5 s (IQR 2.3–3.9 s) in the supine position and 3.2 s (IQR 2.3–3.6 s) in the upright position. BFT is a new metric that provides important information about bolus transit through the EGJ. An assessment of BFT will determine when the EGJ is open and will also provide a useful method to accurately assess trans-EGJ pressure gradients during flow.

Publisher

American Physiological Society

Subject

Physiology (medical),Gastroenterology,Hepatology,Physiology

Cited by 48 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3