Selective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes

Author:

Branner Almut1,Stein Richard B.2,Normann Richard A.1

Affiliation:

1. The Center for Neural Interfaces, Department of Bioengineering, University of Utah, Salt Lake City, Utah 84112; and

2. Division of Neuroscience, University of Alberta, Edmonton, Alberta T6G 2S2, Canada

Abstract

Restoration of motor function to individuals who have had spinal cord injuries or stroke has been hampered by the lack of an interface to the peripheral nervous system. A suitable interface should provide selective stimulation of a large number of individual muscle groups with graded recruitment of force. We have developed a new neural interface, the Utah Slanted Electrode Array (USEA), that was designed to be implanted into peripheral nerves. Its goal is to provide such an interface that could be useful in rehabilitation as well as neuroscience applications. In this study, the stimulation capabilities of the USEA were evaluated in acute experiments in cat sciatic nerve. The recruitment properties and the selectivity of stimulation were examined by determining the target muscles excited by stimulation via each of the 100 electrodes in the array and using force transducers to record the force produced in these muscles. It is shown in the results that groups of up to 15 electrodes were inserted into individual fascicles. Stimulation slightly above threshold was selective to one muscle group for most individual electrodes. At higher currents, co-activation of agonist but not antagonist muscles was observed in some instances. Recruitment curves for the electrode array were broader with twitch thresholds starting at much lower currents than for cuff electrodes. In these experiments, it is also shown that certain combinations of electrode pairs, inserted into an individual fascicle, excite fiber populations with substantial overlap, whereas other pairs appear to address independent populations. We conclude that the USEA permits more selective stimulation at much lower current intensities with more graded recruitment of individual muscles than is achieved by conventional cuff electrodes.

Publisher

American Physiological Society

Subject

Physiology,General Neuroscience

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3