The effect of body mass and sex on the accuracy of respiratory magnetometers for measurement of end-expiratory lung volumes

Author:

Avraam Joanne12,Bourke Rosie2,Trinder John1,Nicholas Christian L.12,Brazzale Danny2,O’Donoghue Fergal J.23,Rochford Peter D.2,Jordan Amy S.12

Affiliation:

1. Melbourne School of Psychological Sciences, University of Melbourne, Melbourne, Australia;

2. Department of Respiratory and Sleep Medicine and Institute for Breathing and Sleep, Austin Health, Heidelberg, Victoria, Australia; and

3. Faculty of Medicine, University of Melbourne, Parkville, Victoria, Australia

Abstract

Respiratory magnetometers are increasingly being used in sleep studies to measure changes in end-expiratory lung volume (EELV), including in obese obstructive sleep apnea patients. Despite this, the accuracy of magnetometers has not been confirmed in obese patients nor compared between sexes. Thus we compared spirometer-measured and magnetometer-estimated lung volume and tidal volume changes during voluntary end-expiratory lung volume changes of 1.5, 1, and 0.5 l above and 0.5 l below functional respiratory capacity in supine normal-weight [body mass index (BMI) < 25 kg/m] and healthy obese (BMI > 30 kg/m) men and women. Two different magnetometer calibration techniques proposed by Banzett et al. [Banzett RB, Mahan ST, Garner DM, Brughera A, Loring SH. J Appl Physiol (1985) 79: 2169–2176, 1995] and Sackner et al. [Sackner MA, Watson H, Belsito AS, Feinerman D, Suarez M, Gonzalez G, Bizousky F, Krieger B. J Appl Physiol (1985) 66: 410–420, 1989] were assessed. Across all groups and target volumes, magnetometers overestimated spirometer-measured EELV by ~65 ml (<0.001) with no difference between techniques (0.07). The Banzett method overestimated the spirometer EELV change in normal-weight women for all target volumes except +0.5 l, whereas no differences between mass or sex groups were observed for the Sackner technique. The variability of breath-to-breath measures of EELV was significantly higher for obese compared with nonobese subjects and was higher for the Sackner than Banzett technique. On the other hand, for tidal volume, both calibration techniques underestimated spirometer measurements (<0.001), with the underestimation being more marked for the Banzett than Sackner technique (0.03), in obese than normal weight (<0.001) and in men than in women (0.003). These results indicate that both body mass and sex affect the accuracy of respiratory magnetometers in measuring EELV and tidal volume.

Funder

National Health and Medical Research Council (Australia)

Australian Research Council

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3