Contact-free unconstraint respiratory measurements with load cells under the bed in awake healthy volunteers: breath-by-breath comparison with pneumotachography

Author:

Isono Shiroh1,Nozaki-Taguchi Natsuko2,Hasegawa Makoto2,Kato Shinichiro2,Todoroki Shinsuke3,Masuda Shigemi3,Iida Norihito4,Nishimura Toshiaki5,Noto Masatoshi4,Sato Yasunori6

Affiliation:

1. Department of Anesthesiology, Graduate School of Medicine, Chiba University, Chiba, Japan

2. Department of Anesthesiology, Chiba University Hospital, Chiba, Japan

3. Functional Technology Department, Section 4, R&D Division Electronic Device and Component, Manufacturing Headquarters and Engineering Headquarters, MinebeaMitsumi, Fukuroi, Japan

4. Sensing Device Business Unit, Electronic Device Division, Electronic Device and Component Manufacturing Headquarters, MinebeaMitsumi, Fujisawa, Japan

5. Sales Planning Office, Bed Sensor Product Management Division, MinebeaMitsumi, Tokyo, Japan

6. Department of Preventive Medicine and Public Health, Keio University School of Medicine, Tokyo, Japan

Abstract

Rate of respiration is a fundamental vital sign. Accuracy and precision of respiratory rate measurements with contact-free load cell sensors under the bed legs were assessed by breath-by-breath comparison with the pneumotachography technique during two different dynamic breathing tasks in 16 awake human adults resting on the bed. The subject voluntarily increased and decreased the respiratory rate between 4 and 16 breaths/min ( n = 8) and 10 and 40 breaths/min ( n = 8) at every 2 breaths in 6 different lying postures such as supine, left lateral, right lateral, and 30, 45, and 60° sitting postures. Reciprocal phase changes of the upper and lower load cell signals accorded with the respiratory phases indicating respiratory-related shifts of the centroid along the long axis of the bed. Bland-Altman analyses revealed 0.66 and 1.59 breaths/min standard deviation differences between the techniques (limits of agreement: −1.22 to 1.36 and −2.96 to 3.30) and 0.07 and 0.17 breaths/min fixed bias differences (accuracy) (confidence interval: 0.04 to 0.10 and 0.12 to 0.22) for the mean respiratory rates of 10.5 ± 3.7 and 24.6 ± 8.9 breaths/min, respectively, regardless of the body postures on the bed. Proportional underestimation by this technique was evident for respiratory rates >40 breaths/min. Sample breath increase up to 10 breaths improved the precision from 1.59 to 0.26 breaths/min. Abnormally faster and slower respirations were accurately detected. We conclude that contact-free unconstraint respiratory rate measurements with load cells under the bed legs are accurate and may serve as a new clinical and investigational tool. NEW & NOTEWORTHY Four load cells placed under the bed legs successfully captured a centroid shift during respiration in human subjects lying on a bed. Breath-by-breath comparison of the breaths covering a wide respiratory rate range by pneumotachography confirmed reliability of the contact-free unconstraint respiratory rate measurements by small standard deviations and biases regardless of body postures. Abnormally faster and slower respirations were accurately detected. This technique should be an asset as a new clinical and investigational tool.

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology