Improved agreement between N2 and SF6 multiple breath washout in healthy infants and toddlers with improved EXHALYZER D® sensor performance

Author:

Sandvik Rikke Mulvad1,Gustafsson Per Magnus23,Lindblad Anders34,Robinson Paul David5,Nielsen Kim Gjerum16

Affiliation:

1. Danish Paediatric Pulmonary Service, Copenhagen University Hospital, Denmark

2. Department of Pediatrics, Central Hospital, Sweden

3. The Sahlgrenska Academy at the University of Gothenburg, Gothenburg, Sweden

4. Gothenburg CF Centre, Queen Silvia Children's Hospital, Sweden

5. Department of Respiratory Medicine, Children's Hospital at Westmead, Australia

6. Department of Clinical Medicine, University of Copenhagen, Denmark

Abstract

Introduction: Recent studies indicate limited utility of nitrogen multiple breath washout (N2MBW) in infancy and advocate for using sulphur hexafluoride (SF6)MBW in this age group. Modern N2MBW systems, such as EXHALYZER D® (ECO MEDICS AG, Duernten, Switzerland), use O2 and CO2 sensors to calculate N2 concentrations (in principle: N2%=100-CO2%-O2%). High O2 and CO2 concentrations have now been shown to significantly suppress signal output from the other sensor, raising apparent N2 concentrations. We examined whether improved Exhalyzer D® N2-signal, accomplished after thorough examination of this CO2 and O2 interaction on gas sensors and its correction, leads to better agreement between N2MBW and SF6MBW in healthy infants and toddlers. Method: Within the same session 52 healthy children aged 1-36 months (mean 1.30 (SD 0.72) years) completed SF6MBW and N2MBW recordings (EXHALYZER D®, SPIROWARE® version 3.2.1) during supine quiet sleep. SF6 and N2 SPIROWARE® files were re-analyzed off-line with in-house software using identical algorithms as in SPIROWARE® with or without application of the new correction factors for N2MBW provided by ECO MEDICS AG. Results Applying the improved N2-signal significantly reduced mean (95% CI) differences between N2- and SF6MBW recorded functional residual capacity (FRC) and lung clearance index (LCI): for FRC, from 26.1 (21.0; 31.2) mL p<0.0001 to 1.18 (-2.3; 4.5) mL p=0.5, and for LCI, from 1.86 (1.68; 2.02) p<0.001 to 0.44 (0.33; 0.55) p<0.001. Conclusion: Correction of N2-signal, for CO2 and O2 interactions on gas sensors resulted in markedly closer agreement between N2MBW and SF6MBW outcomes in healthy infants and toddlers.

Funder

Vertex Pharmaceuticals

Danmarks Lungeforening

Danish Cystic Fibrosis Foundation

The Research Fund at Skaraborg Hospital, Skovde, Sweden

the SAKS Research Foundation, Uppsala, Sweden

Sanatorie laege Ellen Pedersens legat

Kong Christian V og dronning Louises Jubilaeums legat

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3